User Guide

Instruction Manual DO 6
61
% Saturation Calibration Limits Factory calibrated absolute value of 10.0%
and below for 0% point & 50% to 200% for
100% point.
Concentration Calibration Window +/- 40% from the factory default
measurement value. Minimum reading
allowed is 2.00 mg/L (ppm).
Temperature Calibration Window +/- 5°C from factory default measurement
Offset Adjustments (% Saturation) +/- 10.0 of reading in Saturation mode
HOLD function Yes
Auto-Off function Selectable Auto Off function. (20 minutes
after last key press)
Display Customs Single 4 Digit LCD
Inputs BNC for DO & 2.5 mm Phono for
temperature
Operating Range 0 to 50 °C
Power Requirements 4 AAA-sized batteries (included)
Battery Life > 700 hours (Alkaline Batteries)
Dimensions Meter: 14 cm (L) x 7 cm (W) x 3.5 cm (H)
Probe: 115 mm (L) x 12 mm (Dia) with 3-ft
cable.
Probe’s membrane housing: 16 mm (Dia)