User Guide

Instruction Manual DO 6
29
5.4.2 Salinity Setting Adjustment
NOTE: This mode is available only from
the mg/L (ppm) Concentration Set Up
menu.
1) Repeat steps 1 to 5 as in the steps of
Pressure Setting Adjustment.
2) Press the HOLD/ENTER key twice till
the meter displays the Salinity setting
mode “b.SAL”.
3) Press the HOLD/ENTER key to enter
the Salinity setting mode.
4) Use the and keys to enter the
salinity of your solution in parts per
thousand (ppt). Factory default is 0.0.
Salinity values up to 50 ppt can be
entered.
5) Press HOLD/ENTER key to confirm
the value. A confirmation indicator
“CO” will be displayed for 1.5 seconds
before returning to the main group
menu “DPr.7”.
6) Press CAL key to return to the
measurement mode.
Figure 17: Salinity Setting Adjustment Sequence
Display will return back to “DPr.7”
subgroup menu
HOLD
ENTER
HOLD
ENTER
“CO” will display for
1.5 seconds