User Guide

Instruction Manual DO 6
16
4) Press the and keys to adjust the meter reading to the correct
temperature value (i.e. the temperature of the temperature bath)
5) Press the HOLD/ENTER key to confirm temperature calibration and return
to measurement mode.
NOTE:
To exit from Temperature Calibration mode without confirming the calibration,
press CAL instead of HOLD/ENTER.
Temperature calibration is restricted to ± 5°C from the factory default
temperature measurement displayed during calibration.