User Guide
EcoScan CON 5 / TDS 5
9
4. Press ∆
∆∆
∆ (up) or ∇
∇∇
∇ (down) key to scroll the reading until the displayed value
matches your standard solution i.e. 1413 µS/cm.
5. Press ENTER key to confirm calibration and the LCD displays the new set
reading and revert to measurement mode.
6. Alternatively you can abort the new calibrated value by pressing CAL key to
exit from calibration mode.
Important: Meter allows a tolerance range of 30% of its measured reading. You
will not able to scroll through to the desired value if reading is out of its
tolerance.
3.2 TDS Calibration (EcoScan TDS 5)
The calibration procedure for the EcoScan TDS 5 is the same as EcoScan CON
5 (refer to section 3.1 for details).
CA
1395
uS
1413
uS