User Guide

Instruction Manual pH 510/ Ion 510
19
5.3 P3.0 Selecting pH Buffer Standard
5.4 P4.0 Meter Reset
This program lets you select between two standard calibration buffer sets,
depending upon your requirements.
The available standards are USA and NIST.
1. Enter the SETUP mode as described in this section earlier. The
meter automatically goes to program P1.0.
2. Press ¿ or À key to scroll through the programs until you view “bUF
in the upper display and P3.0 in the lower display. The buffer
annunciator also appears.
3. Press ENTER key to access the program P3.0. The lower display will
read either ”USA” or “nSt (depending on the previous setting you
have made; factory default is USA.)
4. Press ¿ or À key to select the buffer set you require:
USA buffers: 4.01, 7.00 and 10.01
nSt buffers: 4.01, 6.86 and 9.18
5. To confirm the buffer standard, press ENTER. All the characters on
the display will flash momentarily and then returns to P3.0.
6. To exit P3.0 program press ¿ or À key to select a new program OR
press the CAL/MEAS to return to the measurement mode.
This program resets and clears all memory data and calibration data
except temperature calibrated value which will be retained. NO is the
default setting.
1. Enter the SETUP mode as described in this section earlier on. The
meter automatically goes to program P1.0.
2. Press ¿ or À key to scroll through the programs until you view rSt in
the upper display and P4.0 in the lower display. The mV and pH
annunciators also appears.
3. Press the ENTER key to access the program P4.0. The lower
display will read NO.
4. Press ¿ or À key to select “NO” or “YES”. Select YES erases all
calibration values for all modes (pH, mV and ion) and all memory
values.
5. To confirm the reset, make sure YES is selected and press ENTER.
All the characters on the display will flash for a second then the
meter returns to the measurement mode.
NOTE
Once YES is selected and ENTER key is pressed; the reset program
clears all old calibration data and memory data. The data is no longer
retrievable. You must calibrate the meter before taking any new
measurements. Meter will however retain buffer set selection (NIST or
USA).