5 SETUP FUNCTIONS
The advanced SETUP mode lets you customize your meter’s preferences and defaults:
P1.0: Memory Clear (Clr)
P2.0: Viewing electrode data (ELE)
P3.0: Selecting buffer sets (bUF)
P4.0: Reset to factory default settings (rSt)
To enter the SETUP mode:
1. Turn meter OFF.
2. With meter off, press and hold the MODE key, and press and release the ON key. The display will show
the [SETUP] indicator. If the display indicates you are in the measurement mode turn the meter off and try
step 2 again until [SETUP] appears.
NOTE: To exit the SETUP mode at anytime without confirming changes press CAL/MEAS until the
measurement mode appears. You may have to press CAL/MEAS key two times before the meter reverts to the
measurement mode.
5.1 P1.0 Clear Memory
Use this program to clear all memory values when you need to store a
new series of values. This lets you avoid confusing the old values with
the new ones. NO is the default setting.
1. Enter the SETUP mode as described above. The meter
automatically goes to program P1.0. Clr appears in the upper
display and P1.0 in the lower display.
2. Press the ENTER key to access the program and the lower display
shows either “nO” or “YES” (depending on last option selected or
factory default i.e. nO).
3. Use ¿ or À key to toggle between “nO” or “YES”
4. Press ENTER key to confirm selection and return to the SETUP
mode. To continue to the next SETUP, press ¿ or À key to select a
new program. Otherwise press the CAL/MEAS key to return to the
measurement mode.