User Guide

Instruction Manual pH 510/ Ion 510
11
3.3 mV Calibration
1. Make sure the meter is on and if necessary, press the MODE key to
select mV mode. The [mV] indicator appears in the upper right-hand
corner of the display.
2. Press the CAL/MEAS key. The [CAL] indicator appears above the
upper display that displays the relative mV reading. The lower
display shows the absolute mV value.
NOTE: If you have never calibrated relative mV or if the meter has
been reset, the value shown in the upper display is the same as the
absolute mV value. The range of adjustment is -1999 mV to 1999
mV (resolution is 1 mV); of which from -199.9 to 199.9 mV the meter
gives you a resolution of 0.1 mV.
3. Press ¿ or À key to enter the relative mV value that matches your
desired reading.
4. Press the ENTER key to confirm the reading and the meter
automatically returns to the measurement mode. The upper display
now shows the relative mV reading. The [R.mV] indicator appears in
the upper right hand corner.
NOTE: New mV calibrations will override existing stored mV
calibration data. The meter retains stored mV calibrations even when
the meter is turned off.