User Guide

Instruction Manual pH 510/ Ion 510
10
If you are performing a 3-point calibration, rinse off the electrode
with deionized water, place it in the next calibration standard and
proceed as follows.
8. Allow meter to stabilize in the next calibration standard. Once the
reading has stabilized, press the ENTER key to confirm the third
calibration point. The upper display flashes for a few seconds then
SLO” appears in the upper display with the mV indicator, the
number in the lower display is the average electrode slope in mV
(see Figure 6). After a few seconds the meter reverts to the
measurement mode. You are now ready to take ion measurement.
NOTE: All the calibration values are successfully stored. Otherwise
error message will appear in the LCD if the calibration was
unsuccessful with no values stored into memory.
NOTE: You may compare the average electrode slope value with
the expected slope value for your electrode from your electrode
manual to verify electrode operation.
HINT: Minimum of 2
points and maximum
of 3 points are
allowed for Ion
Calibration. All
calibration points
should be at least one
decade apart, i.e.
valid calibration
points are 0.1, 1, 10
ppm or 1, 10, 100 ppm
or 0.1, 1 ppm or 1, 10
ppm or 10, 100 ppm.
Figure 6: Slope value in mV
Figure 7: Error message Er3
if slope not within range