User Guide

Instruction Manual pH 510/ Ion 510
7
All new calibrations will override existing stored calibration data at these calibration points. It is recommended
that you perform at least a 2-point calibration using buffers that bracket (one above and one below) the
expected sample range, starting with pH 7 buffer first. You can perform a 1-point calibration, but make sure that
the buffer value is close to the sample value you are measuring. Select a pH 4, 7 or 10 buffer in sachets or
bottles from standard buffer solutions.
1. Press ON/OFF key. All the LCD segments will be displayed
momentarily. The LCD switches to pH measurement mode. If
necessary press MODE key to select pH mode.
NOTE: The meter starts in the same measurement mode that it
was switched off.
2. The customized LCD display will indicate the following:
Display Remarks
MEAS Measurement mode is selected
pH Unit of measurement
o
C Measured temperature reading
ATC Automatic Temperature Compensation
(If temperature probe or “All-in-One” probe is
connected)”
3. Rinse the pH electrode well with deionized water or rinse solution.
If using the ATC function with a separate temperature probe,
thoroughly rinse the temperature probe as well. (Do not wipe the
electrode or the temperature probe as this may cause a build-
up of electrostatic charge and will create calibration and
measurement instability!).
One Point Calibration
4. Select the first pH buffer, say pH 7.00 buffer and pour some into a
clean container.
5. Dip both probes into the standard calibration buffer. The end of
the probe must be completely immersed into the buffer. Stir the
probes gently to create a homogenous solution.
6. Press CAL/MEAS key to enter the pH calibration mode. The
annunciator [CAL] appears on top of the LCD to indicate the
meter is in pH calibration mode. The upper display will show the
measured reading while the lower display indicates appropriate
pH standard buffer solution being used.
NOTE: The meter automatically recognizes the buffers in the
buffer standard you have set in the SETUP mode, i.e. either USA
(pH 4.01, 7.00 or 10.01) or NIST (pH 4.01, 6.86 or 9.18) buffers.
Figure 1: 1st-point calibration