f. Repeat the above step c through e for the second and subsequent
calibration points. Meter automatically returns to the Measure mode
after the last 5
th
calibration point has been performed. In the pH
Measure mode, the screen shows all the buffer values being
standardized. An error message Err2 will be displayed if there is more
than 6 points pH standardization being performed.
Note:
If the “EURO” buffer is selected, meter returns to measurement mode
automatically after 6
th
point calibration.