User Guide

Instruction Manual CyberScan pH1500
15
4.1.3 P3.0 pH Buffer Group
This setup option allows you to select from 3 different buffer groups,
each containing 5 buffers, for auto buffer recognition.
The 3 buffer groups are:
USA buffers: 2, 4, 7, 10 and 12
European buffers: 1, 3, 6, 8 and 10
NIST buffers: 1.68, 4.01, 6.86, 9.18 and 12.46
To Set pH Buffer Group
1. Access the Select Buffer menu from the pH Setup Screen by
pressing the setup key.

press to set value
press to accept
enter
ELECTRODE
BUFFER
select
2. Press and keys to scroll until display shown as above.
save
view
.