User Guide
Instruction Manual CyberScan pH1500
15
4.1.3 P3.0 pH Buffer Group
This setup option allows you to select from 3 different buffer groups,
each containing 5 buffers, for auto buffer recognition.
The 3 buffer groups are:
USA buffers: 2, 4, 7, 10 and 12
European buffers: 1, 3, 6, 8 and 10
NIST buffers: 1.68, 4.01, 6.86, 9.18 and 12.46
To Set pH Buffer Group
1. Access the Select Buffer menu from the pH Setup Screen by
pressing the setup key.
press to set value
press to accept
enter
ELECTRODE
BUFFER
select
2. Press ▲ and ▼ keys to scroll until display shown as above.
save
view
.