3.2 Temperature Probe
For more accurate pH measurement, a temperature probe (supplied
with the meter) is used for Automatic Temperature Compensation
(ATC). Once the temperature probe is connected into its socket
(located at the back of meter), the ATC icon is displayed on the
screen to indicate that the meter is in ATC Measure mode. If the
temperature probe is not connected the meter displays 25.0
o
C or 77
o
F as the default temperature value. When in ATC mode, the meter
allows fine offset temperature adjustment of ±5.0
o
C (±9.0
o
F) from
mV mode by pressing the std key. Use or key to set the desired
value and press enter key to confirm. The meter returns to the
Measure mode.