32
9.3 Program 3 –Meter Configuration
The following three options are available to the user for
customization:
9.3.1 P3.0 : Resolution
The resolution of the instrument in the pH measurement
mode can be user selected to 0.1 or 0.01 pH. Under
default condition, the instrument is set for 0.01 pH
resolution.
LCD Value Description
0.01 A resolution of 0.01 pH is shown.
0.1 A resolution of 0.1 pH is shown.
1. Use MI/
!
or MR/
"
keys to select the resolution
desired. See Figure 33.
2. Press CON key to continue to P3.1.
9.3.2 P3.1 : Ready selection
Activation of the READY option ensures that the READY
indicator is displayed when the electrode reading has
stabilized.
The stabilization tolerances for the various measurement
ranges are as indicated below. Under default conditions,
i.e. when the [SETUP] mode is not programmed, the
instrument has this option activated.
MODE VARIATION RANGE
pH ± 0.01 pH
mV ± 0.8 pH < 400 mV
mV ± 1.2 mV > 400 mV
ON READY Indicator will appear
when the reading stabilizes
within ± 0.01 pH
OFF READY indicator will not appear.
Figure 33: Setting pH resolution