14
4.3 pH Calibration (without ATC)
When the CyberScan meter is not used with a temperature probe, it is necessary to calibrate
the temperature first prior to pH calibration.
1. Calibrate for temperature as described in the procedure in Section 4.4.1 -
Temperature
Calibration Without ATC Probe
.
2. Next, proceed with the pH calibration as described in the above section.
3. The meter is now calibrated at a fixed temperature.
NOTE: To exit this program without confirming the temperature calibration value, DO NOT
press CON. Press CAL/MEAS instead.