Instruction manual

16
6. Press the ENTER key to confirm temperature
calibration. The CON indicator will flash to
confirm calibration.
See Figure 20.
NOTES
To exit from Temperature Calibration mode without
confirming calibration, DO NOT press ENTER in
step 6. Press CAL/MEAS instead.
Temperature calibration is restricted to ±5°C from
the initial value displayed during calibration (shown in the secondary display).
Figure 20 – CONFIRM temperature
calibration