Instruction manual
14
8. Press the ENTER key to confirm calibration.
Upon confirmation, the CON indicator appears
briefly. The meter automatically switches back into
Measurement mode. The display now shows the
calibrated, temperature compensated conductivity
value.
See Figure 16.
9. For calibration in other ranges (maximum:
four ranges) repeat steps 1 through 9 with the
appropriate calibration standards. See page
24 for our selection of high-quality Eutech
Instruments conductivity calibration solutions.
NOTES:
To exit from Conductivity Calibration mode without confirming calibration, DO NOT press
ENTER in step 8. Press CAL/MEAS instead.
If the calibration value input into the meter is different from the initial value displayed by more
than 20%, the ERR annunciator appears in the lower left corner of the display.
Figure 16 - CONFIRM the calibration
value