Instruction manual

9
4.2. Preparing the Meter for Calibration
The pH/conductivity/temperature probe included with this meter is designed for use with this
meter only. Do not substitute other types of probes or electrodes. For a replacement probe,
see the “Accessories” section, pages 23 - 24.
Be sure to remove the protective rubber cap of the probe before calibration or
measurement. Immerse the probe in tap water for 10 minutes before calibrating or taking
readings to saturate the pH electrode surface and minimize drift.
Wash your probe in deionized water after use, and store in pH 4.0 or 7.0 electrode storage
solution.
Do not reuse buffer solutions after calibration. Contaminants in the solution can affect the
calibration, and eventually the accuracy of the measurements.
4.3. pH Calibration
This instrument is capable of up to 3-point pH calibration to ensure accuracy across the entire
pH range of the meter. You can perform 1, 2, or 3 point calibration with standard pH buffers
of 4.00, 7.00 and 10.00.
We recommend that you perform at least 2-point calibration using standard buffers that
bracket (one above and one below) the expected sample range. You can also perform a 1-
point calibration, but make sure that the buffer value is close to the sample value you are
measuring.
This meter features three pre-programmed pH buffers (pH 4.00, 7.00 and 10.00). The meter
automatically recognizes and calibrates to these standard buffer values, which makes pH
calibration faster and easier. See page 24 for information on our high-quality Eutech
Instruments pH buffers.
Calibrating for pH:
1. If necessary, press the MODE key to select pH mode. The pH indicator appears in the
upper right hand corner of the display.
2. Rinse the probe thoroughly with de-ionized water or a rinse solution. Do not wipe the
probe; this causes a build-up of electrostatic charge on the glass surface.