5.6 Set Stability Criteria
Current MODE
Select
D.O. Stability Criteria
AUTO
S
T
enter
Fast
Slow
Medium
This setup screen allows you to choose how quickly the meter will
accept a reading as stable.
There are three settings: fast, medium, and slow.
To set Stability Criteria
1. Access the Set Stability Criteria screen from the DO/ BOD/ OUR/
SOUR Setup screen. The current stability setting is displayed on
the screen.
2. Use the arrow keys to highlight the parameter you wish to
change.
3. Touch enter to accept the stability criteria and return to the Setup
screen.
OR
Touch exit to return to the Setup screen, without making any
changes.