44
Set SOUR Configuration
This option allows the user select parameters that define the criteria
for the Specific Oxygen Uptake Rate (OUR) test. These include
Dilution Factor, Minimum Time (minutes) duration of the test,
Maximum Time duration, Minimum Starting DO, Minimum Ending
DO, and Solids Weight (g/L). This setup option appears only when
the meter is in the SOUR mode.
1. Access the Set SOUR Configuration from the Setup screen while
in SOUR mode.
2. Use the arrow keys to highlight the parameter you wish to
change.
3. Touch edit.
4. Use the numeric keypad that appears to key in the value desired
for the chosen parameter.
5. Touch enter to accept the value.