5.5 Set Configurations for BOD/ OUR and SOUR
(Not applicable in D.O. mode)
Set BOD Configuration
This option allows the user to select certain parameters that define
the criteria for seed and sample acceptance. Theses parameters are
Seed Minimum Data DO (DO
initial
- DO
final
). Seed minimum endpoint
DO, Sample minimum Delta DO, and sample Minimum endpoint DO.
These parameters will aid the user to disregard “bad” samples. This
setup option appears only when the meter is in the BOD mode.
1. Access the Set BOD Configuration from the Setup screen while
on BOD mode.
2. Use the arrow keys to highlight to parameter you wish to change.
3. Touch edit.
4. Use the numeric keypad that appears to key in the value desired
for the chosen parameter.
5. Touch enter to accept the value.
Set OUR Configuration
This option allows the user to select parameters that define the
criteria for the Oxygen Uptake Rate (OUR) test. These include
Dilution Factor, Minimum Time (minutes) duration of the test,
Maximum Time duration, Minimum Starting DO, and Minimum Ending
DO. This setup option appears only when the meter is in the OUR
mode.
1. Access the Set OUR Configuration from Setup screen while in
OUR mode.
2. Use the arrow keys to highlight to parameter you wish to change.
3. Touch edit.
4. Use the numeric keypad that appears to key in the value desired
for the chosen parameter.
5. Touch enter to accept the value.