Instruction Manual CyberScan DO 5000 Bench Dissolved Oxygen Meter (Also Applicable for CyberScan PCD 5500) Technology Made Easy ... 68X292330 Rev.
PREFACE Thank you for selecting the Eutech Instruments CyberScan DO 5000 bench meter. This meter measures dissolved oxygen and temperature (°C/ °F). This instruction manual serves to explain the use of the CyberScan DO 5000 bench meter as a step-by-step operational guide to help you familiarize with the meter’s features and functions. It is structured sequentially with illustration of diagrams that explains the various functions and setup menus available.
TABLE OF CONTENTS 1 INTRODUCTION 2 UNPACKING THE METER 3 GETTING STARTED 1.1 3.1 3.2 3.3 3.4 4 Introducing the CyberScan Series Connectors Dissolved Oxygen Probe Touch Screen Operation Button functions SYSTEM SETUP 4.1 Set Date 4.2 Set Time 4.3 Set Beeper Status 4.4 Set Print Configuration 4.4.1 Set Baud Rate 4.4.2 Set Number of Bits 4.4.3 Set Stop Bits 4.4.4 Set Parity 4.4.5 Set Serial Output Mode 4.5 Set Barcode Configuration 4.6 Set Operator 4.7 Set Display Contrast 4.
Instruction Manual CyberScan DO5000 1 1.1 INTRODUCTION Introducing the CyberScan Series Thank you for selecting a Eutech Instruments CyberScan Bench meter. This manual describes the operation of the CyberScan DO 5000 bench meter. The state-of-art meter that you have purchased is easy to operate and will guide you through the various functions by displaying easy to understand prompts.
Instruction Manual CyberScan DO5000 2
Instruction Manual CyberScan DO5000 2 UNPACKING THE METER The following is a listing of what you should have received with your new CyberScan DO 5000 meter. Meter with kit includes Meter Power adapter (120 VAC/ 12VDC) OR (220 VAC/ 12 VDC) depending on order code. BOD Probe Instruction Manual If any of these items are missing, please contact your nearest Authorized Distributor.
Instruction Manual CyberScan DO5000 3 3.1 GETTING STARTED Connectors 1. Review the layout and arrangement of the rear connector panel. 2. Connect the power cable to the rear connector panel power jack and to a powers source.
Instruction Manual CyberScan DO5000 3.2 Dissolved Oxygen Probe This meter is equipped to use the BOD probe. 1. Carefully review the instruction sheet accompanying the probe for setup and conditioning information. 2. Connect the probe to the meter by plugging it into the DIN connector on the back of the meter. Permit the probe to warm up for 30 minutes. 3. Store the DO probe in a BOD bottle filled with 1 inch of water.
Instruction Manual CyberScan DO5000 1. Connect the electrode arm to the base. 2. Connect the power cable to the connector cable to the rear connector panel power jack and to a power source. To connect RS232, see Data Management on Section 9 below, page 93.
Instruction Manual CyberScan DO5000 Standby screen EUTECH INSTRUMENTS 11:11 am Touch anywhere to resume 7
Instruction Manual CyberScan DO5000 Meter Main Screen EUTECH INSTRUMENTS CyberScan DO 5000 DO BOD OUR Select from the options to the right SOUR system 8
Instruction Manual CyberScan DO5000 3.3 Touch Screen Operation The CyberScan DO 5000 meter operates with a state of art touch screen. The touch screen makes this the easiest meter on the market to operate and care for. When this meter is first plugged in, the STANDBY screen will appear. Touch anywhere on this screen to access the functions of the meter. The buttons on the right side of the screen control all of the functions of the meter.
Instruction Manual CyberScan DO5000 3.4 Button functions The touch screen of your CyberScan research DO meter has “buttons” along the right side of the screen that are common to many of the screens. The following indicates the function of these common buttons. This is the standby button and it allows you to access the standby mode. When in standby, the meter will not take measurements. It is in a state of rest.
Instruction Manual CyberScan DO5000 std This button accesses the standardization screen from the various measurement modes and initiates standardization of the meter once the standardization screen is accessed. meas This button is the measure button and directs the meter to measure your sample when in the Auto Read mode. setup This button will access he setup screens for the measuring mode that you are currently using.
Instruction Manual CyberScan DO5000 4 SYSTEM SETUP System Setup SYSTEM SETUP OPTIONS S - Set Date T - Set Time - Set Beeper Status - Set Print Configuration - Set Barcode Configuration - Set Operator - Set Display Contrast - Display Meter Information - Reset to Factory Defaults 12 edit
Instruction Manual CyberScan DO5000 EUTECH DO INSTRUMENTS CyberScan DO 5000 BOD OUR Select from the options to the right SOUR system January 15, 2002 11:11 am System Setup SYSTEM SETUP OPTIONS S - Set Date T - Set Time - Set Beeper Status - Set Print Configuration - Set Barcode Configuration - Set Operator - Set Display Contrast - Display Meter Information - Reset to Factory Defaults 13 edit
Instruction Manual CyberScan DO5000 Access system setup System setup can be accessed from the Setup screen (Page 12). The system setup function allows you to customise the meter display options to meet your personal preference. Once set, these will rarely need to be changed. To Access System Setup 1. Touch anywhere on the standby screen 2. Touch system on the setup screen The system setup options are now displayed on the screen. To Access a System Setup Option 1.
Instruction Manual CyberScan DO5000 4.
Instruction Manual CyberScan DO5000 This screen can be used to set the present date which will be displayed on the measure screens. This date will also be printed on demand and stored in the data storage center of the meter when data is saved. There are two format options for the date: month/ day/ year (M/ D/ Y) or the European format of day/ month/ year (D/ M/ Y) To set date 1. Access the Set Date screen from the System Setup screen. The current date and numeric keypad are displayed on the screen. 2.
Instruction Manual CyberScan DO5000 4.
Instruction Manual CyberScan DO5000 This screen can be used to set the present time which will be displayed on the measure screens. This time will also be printed on demand and stored in the data storage center of the meter when data is saved. There are two formats options for the time. The clock can be set as either a 12 hour clock or a 24 hour clock. To set time 1. Access the set time screen from the System Setup screen. The current time and numeric keypad are displayed on the screen. 2.
Instruction Manual CyberScan DO5000 4.
Instruction Manual CyberScan DO5000 This screen allows you to turn on or off the beeper. You may choose to have an audible signal when the meter recognises that the current measurement is stable, each time a function button is touched and/ or when the set limits of a measurement mode have been exceeded. To Set Beeper Status 1. Access the set beeper status screen from the System Setup screen. The current beeper status is displayed on the screen. 2.
Instruction Manual CyberScan DO5000 4.
Instruction Manual CyberScan DO5000 You can adjust the print configuration of this meter from this screen. The configuration of the following screens must match the configuration of the printer or computer to which the data will be sent. To Set Print Configuration 1. Access the Print Configuration screen from the System Setup screen. The current Print Configuration is displayed on the screen. 2. Use the arrow keys to highlight the configuration option to be modified. 3.
Instruction Manual CyberScan DO5000 4.4.1 Set Baud Rate Set Baud Rate 300 Current BAUD RATE 110 S T 300 600 1200 2400 enter 4800 9600 19200 38400 This configuration option will control the speed at which the data will be transmitted by the printer. This parameter needs to match the baud rate designated by the printer or computer. To Set Baud Rate 1. Access the Set Baud Rate screen from the Set Print Configuration screen. The current baud rate is displayed on the screen. 2.
Instruction Manual CyberScan DO5000 4.4.2 Set Number of Bits Set Number of Bits enter Current NUMBER OF BITS 7 8 To Set Number of Bits 1. Access the Set number of Bits screen from the Set Print Configuration screen. The current number of bits is displayed on the screen. 2. Touch 7 or 8 to select the number of bits. 3. Touch enter to accept the bit value and return to the Set Print Configuration screen. OR Touch exit to return to the Set Print Configuration screen, without making any changes.
Instruction Manual CyberScan DO5000 4.4.3 Set Stop Bits Set Stop Bits enter 1 Current STOP BITS 2 To Set Stop Bits 1. Access the Set Stop Bits screen from the Set Print Configuration screen. The current number of bits is displayed on the screen. 2. Touch 1 or 2 to set the desired number of stop bits. 3. Touch enter to accept the stop bit value and return to the Set Print Configuration screen. OR Touch exit to return to the Set Print Configuration screen, without making any changes.
Instruction Manual CyberScan DO5000 4.4.4 Set Parity Set Parity Current PARITY enter NONE EVEN ODD To Set Parity 1. Access the Set Parity screen from the Set Print Configuration screen. The current Parity is displayed on the screen. 2. Touch ODD or EVEN or NONE to set the desired parity. 3. Touch enter to accept the parity setting and return to the Set Print Configuration screen. OR Touch exit to return to the Set Print Configuration screen, without making any changes.
Instruction Manual CyberScan DO5000 4.4.5 Set Serial Output Mode Set Serial Output Mode enter TEXT Current MODE DATA To Set Serial Output Mode 1. Access the serial output mode from the set print configuration screen. The current mode is displayed. 2. Touch Data or Text on your selection 3. Touch enter to accept the choice and return to the Set Print Configuration screen.
Instruction Manual CyberScan DO5000 4.5 Set Barcode Configuration Set Barcode Configuration enter OFF Current MODE ON Use this option, as well as Set Print Configuration, to prepare the meter for use with a barcode scanner. Set to “ON” is using a barcode scanner. To Set Barcode Configuration 1. Access the Set Barcode Configuration screen from the System Setup screen. The current mode is displayed. 2. Touch ON or OFF on your selection 3. Touch enter to accept the mode.
Instruction Manual CyberScan DO5000 4.6 Set Operator Set Operator enter Current Operator A B C 1 D E F 2 G H I 3 J K L 4 M N O 5 P Q R 6 S T U 7 V W X 8 Y Z - 9 BS / .
Instruction Manual CyberScan DO5000 This option allows you to identify the user of the meter. This information can be saved in the meter’s memory. It can also be printed out with measurement data on demand. The operator identification can be up to 9 characters. To Set Operator 1. Access the Set Operator screen from the System Setup screen. The current operator identification is displayed on the screen. 2. Touch clear to remove the current operator identification. 3.
Instruction Manual CyberScan DO5000 4.
Instruction Manual CyberScan DO5000 This option allows you to change the contrast on the screen to improve the readability of the information presented on the screen. The numbering system that appears on the screen is from 0 to 25. The darkest setting is 0 and the lightest setting is 25. To Set Display Contrast 1. Access the Set Display Contrast screen from the System Setup screen. The current display contrast value is displayed on the screen. 2.
Instruction Manual CyberScan DO5000 4.8 Display Meter Information EUTECH INSTRUMENTS CyberScan DO 5000 Unit Serial Number: Software Revision: XX999XXX 1.04b This screen displays the model number, serial number and current software revision of your meter.
Instruction Manual CyberScan DO5000 4.9 Reset to Factory Defaults Reset to Factory Defaults You are about to reset all parameters of the meter to factory default settings. Are you sure you want to do this? YES NO This screen allows you to reset all functions and setup parameters of the meter of the settings originally programmed at the factory. To Reset to Factory Defaults 1. Access the Reset to Factory Defaults screen from the System Setup screen. 2.
Instruction Manual CyberScan DO5000 5 D.O. / BOD/ OUR/ SOUR SETUP D.O. Setup D.O.
Instruction Manual CyberScan DO5000 5.1 To Access Setup From the DO, BOD, OUR and SOUR screens 1. Touch DO, BOD, OUR, or SOUR from the main screen to access any of these modes. 2. Touch setup 3. Use the arrow keys to highlight the setup option that you would like to review 4. Touch edit to access the screen for the selected option.
Instruction Manual CyberScan DO5000 5.2 Set Sample ID# Set D.O. Sample ID# enter Manual ID# A B C 1 D E F 2 G H I 3 J K L 4 M N O 5 P Q R 6 S T U 7 V W X 8 Y Z - 9 BS / .
Instruction Manual CyberScan DO5000 When this option is active, each time you touch print on the Measure Screen the DO values along with date/time and the sample ID will be sent to data storage. You can manually enter an alphanumeric identification of up to 9 characters for any sample or you can have the meter sequentially number your samples beginning at the number of your choice. You may also choose to deactivate the sample ID. To Set Sample ID Manual ID Assignment 1.
Instruction Manual CyberScan DO5000 Set D.O. Sample ID# enter Sequential ID# 1 2 3 4 5 6 24hr clear 7 8 9 BS 0 .
Instruction Manual CyberScan DO5000 Sequential ID# Assignment 1. Access the Set Sample ID# screen from the Setup screen. 2. Touch seq for sequential ID# assignment. The current ID# is displayed on the screen. 3. Touch clear to delete the current ID#. 4. Use the alphanumeric keypad on the screen to enter the number that you would like your sequential ID# assignment to begin with. Every time you touch print on the measure screen, the ID# will increase by 1.
Instruction Manual CyberScan DO5000 5.3 Select Calibration Mode Select D.O. Calibration Mode enter Current MODE AUTO MAN This mode permits a choice between automatic and manual standardization. In most case, automatic calibration is preferred. However, if the standard is not saturated and the oxygen concentration is known, (by titration, for example), manual calibration offers the user capability of entering the known oxygen concentration via a numeric keypad. 1.
Instruction Manual CyberScan DO5000 5.4 Select Auto Read Mode (Only available in D.O. mode) Select D.O. Auto Read Mode enter Current MODE AUTO MAN You can use this meter when the Auto Read Mode function is active or when it is inactive. When the Auto Read function is active, the meter will lock onto a reading when the meter recognises it as stable. The meter will not deviate from this reading until meas is touched.
Instruction Manual CyberScan DO5000 5.5 Set Configurations for BOD/ OUR and SOUR (Not applicable in D.O. mode) Set BOD Configuration This option allows the user to select certain parameters that define the criteria for seed and sample acceptance. Theses parameters are Seed Minimum Data DO (DOinitial- DOfinal). Seed minimum endpoint DO, Sample minimum Delta DO, and sample Minimum endpoint DO. These parameters will aid the user to disregard “bad” samples.
Instruction Manual CyberScan DO5000 Set SOUR Configuration This option allows the user select parameters that define the criteria for the Specific Oxygen Uptake Rate (OUR) test. These include Dilution Factor, Minimum Time (minutes) duration of the test, Maximum Time duration, Minimum Starting DO, Minimum Ending DO, and Solids Weight (g/L). This setup option appears only when the meter is in the SOUR mode. 1. Access the Set SOUR Configuration from the Setup screen while in SOUR mode. 2.
Instruction Manual CyberScan DO5000 5.6 Set Stability Criteria Select D.O. Stability Criteria Current MODE AUTO S Fast Medium T Slow enter This setup screen allows you to choose how quickly the meter will accept a reading as stable. There are three settings: fast, medium, and slow. To set Stability Criteria 1. Access the Set Stability Criteria screen from the DO/ BOD/ OUR/ SOUR Setup screen. The current stability setting is displayed on the screen. 2.
Instruction Manual CyberScan DO5000 5.7 Set Temperature Units Select D.O. Temperature Units enter C Current UNITS F K 1. Access the Select Temperature Units screen from the DO/ BOD/ OUR/ SOUR Setup screen. The current temperature unit is displayed on the screen. 2. Use the arrow keys to highlight the parameter you wish to change. 3. Touch enter to accept the temperature unit and return to the Setup screen. OR Touch exit to return to the Setup screen, without making a change.
Instruction Manual CyberScan DO5000 5.8 Set Measurement Units Select D.O. Measurement Units Current UNITS %sat S mg/L %sat T mbar enter 1. Access the Select Measurement Units screen from the DO/ BOD/ OUR/ SOUR Setup screen. 2. Use the arrow keys to highlight desired unit. 3. Touch enter to accept the measurement unit and return to the Setup screen. OR Touch exit to return to the Setup screen, without making a change.
Instruction Manual CyberScan DO5000 5.9 Set the Salinity Value of Your Sample 1. Access the Set Salinity screen from the DO/ BOD/ OUR/ SOUR Setup screen 2. Touch clear to delete the current value 3. Key in the desired new value using the numeric keypad 4. Touch enter to accept the new value and return to the Setup screen Set D.O Salinity enter Current Salinity 3.2 ppt 1 2 3 4 5 6 7 8 9 0 .
Instruction Manual CyberScan DO5000 5.10 Set Barometer Set Barometer enter Current Pressure 668.5 mmHg 1 2 3 4 5 6 7 8 9 0 .
Instruction Manual CyberScan DO5000 The DO 5000 has an internal barometer. When it is no longer reflects the true barometer pressure, it must be calibrated. The Set Barometer option allows you to do this. To Set Barometer 1. Access the Set Barometer screen from the DO/ BOD/ OUR/ SOUR setup screen. The current pressure is displayed on the screen 2. Choose the pressure units desired by touching one of the two unit buttons, mm Hg or mbars. 3.
Instruction Manual CyberScan DO5000 5.11 Set Alarm Limits Set D.O. Limits S Current LIMITS - D.O. Alarm OFF - D.O. Minimum 0.00 - D.O.
Instruction Manual CyberScan DO5000 This option allows you to set alarm limits for the DO/ BOD/ OUR/ SOUR measuring mode. If the DO value of the measurement is outside the boundaries set by the minimum and maximum limits, an audible alarm and/ or a visual warning will appear to let you know that your sample measurement was outside of the set limits. To Set Alarm Limits 1. Access the Set Alarm Limits screen from the DO/ BOD/ OUR/ SOUR Setup screen. The current alarm limits are displayed on the screen. 2.
Instruction Manual CyberScan DO5000 5.12 Set Print Criteria Set D.O.
Instruction Manual CyberScan DO5000 This screen allows you to select which criteria are printed with the measurement when you print the data or send it to the computer. The status of the current print criteria is displayed on the screen. The criteria option is active if “ON” appears to the right of the option. It is inactive if “OFF” appears to the right of the option. Any active criteria will be printed on demand. To Set Print Criteria 1.
Instruction Manual CyberScan DO5000 5.13 Set Print Interval Set D.O. Print Interval enter Current INTERVAL manual 1 2 3 4 5 6 stable man 7 8 9 timed BS 0 / Use numeric touchpad to input the desired print interval in seconds and then touch enter to accept Touch stable to print upon stable readings Touch man to select manual printing interval You have three options for setting the print interval: manual printing, stable reading printing, and timed interval printing.
Instruction Manual CyberScan DO5000 2. Touch MAN to set the meter for manual printing. Touch enter to accept the print interval mode and return to the Setup screen. Printing is now done manually by touching print on the Measure screen. OR Touch exit to return to the Setup screen, without making any changes. For Stable Reading Printed In this mode, data is printed every time the meter recognises the current measurement as stable. 1.
Instruction Manual CyberScan DO5000 5.14 Set Data Storage Criteria Set D.O.
Instruction Manual CyberScan DO5000 This screen allows you to select what criteria are stored in the meter’s memory with the measurement when you save the data. Data is stored only if a Sample ID has been assigned. The status of the current data storage criteria is displayed on the screen. The criteria option is active if “ON” appears to the right on the screen. It is inactive if “OFF” appears to the right of the option. All storage criteria will be stored in the meter’s memory with the measurement.
Instruction Manual CyberScan DO5000 5.15 Set Display Resolution Set D.O.
Instruction Manual CyberScan DO5000 Here you can set the display resolution on the screen. Choose high for two decimal points in mg/L, and for one decimal point in % sat and mbar. Choose low for one decimal point in mg/L and one unit resolution in % sat and mbar. To Set Display Resolution 1. Access the Set Display resolution screen from the DO/ BOD/ OUR/ SOUR Setup screen. 2. The current Display Resolution is displayed on the screen. 3. Touch high or low to select the desired resolution. 4.
Instruction Manual CyberScan DO5000 5.16 Set Display Configuration Set D.O.
Instruction Manual CyberScan DO5000 This particular function will allow you to choose what information you would like to be displayed on the DO/ BOD/ OUR/ SOUR Measure screen, particularly the information contained in the data box at the bottom of that screen. To Set Display Configuration 1. Access the Set Display Configuration screen from the DO/ BOD/ OUR/ SOUR Setup screen. 2. Use the arrow keys to highlight the item you want to change. 3. Touch ON or OFF to change the status of the criteria 4.
Instruction Manual CyberScan DO5000 5.
Instruction Manual CyberScan DO5000 The meter has memory capacity of up to 250 data points. The view stored data screen allows you to sort and look at specific data points. The stored data can be sorted by sample identification number, date or operator identification number. To View Stored Data 1. Access the view stored data screen from the DO/ BOD/ OUR/ SOUR Setup screen. The number of data points in the memory and the sorting options are now displayed on the screen. 2.
Instruction Manual CyberScan DO5000 To sort by Date 1. Access the Date sort option from the View Stored Data screen. 2. Touch clear to delete the current date. 3. Use the numeric keypad to enter the date on which the data points you want to view were saved. Be sure to use/ to separate the month, the day and the year. 4. Touch enter. All data will be sorted by the meter and the first data point displayed on the screen will be the most recent data point saved under the selected Date. 5.
Instruction Manual CyberScan DO5000 If an operator ID is entered and no data points are stored with that operator ID, you will see a message indicating the operator ID was not found. Touch OK to return to the operator ID keypad and enter a new operator ID. Note: Even if you do not know the appropriate information to access a specific data point, you can access the stored data through any of the sort options. Highlight the sort option of interest and touch enter to access the sort screen.
Instruction Manual CyberScan DO5000 6 D.O. OPERATION Measure std mg/L 8.21 meas STABLE STND Not Standardized setup Touch meas to measure sample or Touch std to access standardize mode August 12, 2002 ID# auto cal auto read ON OFF 9.32am ATC 25.3°C Salinity 0.
Instruction Manual CyberScan DO5000 6.1 DO Standardization The CyberScan DO 5000 has two standardization modes. There is auto- standardization, merely requiring the touch of a button to complete. There is also manual standardization. Here the DO value, in either % Saturation or mg/L units, of the standard will be entered manually by means of a keypad, which appears on the screen. It is necessary to standardize the meter when the probe is in an environment with a known concentration of oxygen.
Instruction Manual CyberScan DO5000 Standardize std mg/L 8.21 meas STABLE 97.9 %sat 8.15 mg/L STND Last std: Aug 12 @ 8:16 am Touch clear to delete previous standard Insert electrode into standard and stir Touch std to standardize mode August 12, 2002 ID# auto cal auto read 9.32am ABC ATC 25.3°C ON Salinity 0.
Instruction Manual CyberScan DO5000 To Standardize the meter in the Automatic Standardization Mode 1. Insert the DO probe into a BOD bottle containing about 1 inch of water. 2. Touch STD on the DO or BOD Measure screen to access the standardized screen 3. Touch clear to delete a previous standardization. OR If the screen says “Not standardized” proceed to step 4. 4. Touch std again to standardize the meter using this buffer. The word MEASURING will flash until the signal is stable.
Instruction Manual CyberScan DO5000 Standardize std mg/L 8.21 meas STABLE 8.15 mg/L 97.9 %Sat STND setup Last std: Aug 12 @ 8:16 am Touch clear to delete previous standard Insert electrode into standard and stir Touch std to standardize mode August 12, 2002 9.32am 25.3°C ATC ON Salinity 0.0 ppt OFF mmHg 732 ID# auto cal auto read Measure std mg/L 8.21 mg/L STABLE 1 STND meas 2 3 4 5 6 7 8 9 Not Standardized Touch meas to measure sample or - 0 .
Instruction Manual CyberScan DO5000 To Standardize the meter in the Manual Standardization Mode 1. Insert the DO into your choice of standard. 2. Touch std on the DO or BOD Measure screen to access the standardized screen 3. Touch clear to delete a previous standardization. OR If the screen says “Not standardized” proceed to step 4. 4. Touch std again to initiate standardization 5. Using the touch pad, input the value of the DO standard in % Saturation, or mg/L. 6.
Instruction Manual CyberScan DO5000 Measure std mg/L 8.21 meas STABLE STND Not Standardized setup Touch meas to measure sample or Touch std to access standardize mode August 12, 2002 ID# auto cal auto read ON OFF 9.32am ATC 25.3°C Salinity 0.
Instruction Manual CyberScan DO5000 Measuring DO with Auto Read ON The measure screen provides a readout of the current sample measurement. If Auto Read is active, the meter will lock onto a reading until meas is touched. If Auto Read is inactive, the meter will continuously monitor the DO and the reading will indicate any change in the DO level, Regardless of the status of the Auto Read mode, a STABLE message is displayed when the measurement meets the stability criteria chosen.
Instruction Manual CyberScan DO5000 EUTECH INSTRUMENTS CyberScan DO 5000 DO BOD OUR Select from the options to the right SOUR system January 15, 2002 11:11 am Measure std mg/L 8.21 meas STABLE STND Not Standardized setup Touch meas to measure sample or Touch std to access standardize mode August 12, 2002 ID# auto cal auto read ON OFF 9.32am ATC 25.3°C Salinity 0.
Instruction Manual CyberScan DO5000 6.2 Biological Oxygen Demand (BOD) Measurement Measure - BOD std mg/L 8.21 STABLE 0:00 STND start Not Standardized setup Touch start to begin measurement or Touch std to access standardize mode August 12, 2002 ID# autocal 9.32am 25.3°C 0.0 ppt 732 ATC ON salinity mmHg To measure BOD, a sample is seeded and incubated in the dark at 20°C for five days.
Instruction Manual CyberScan DO5000 With the CyberScan DO 5000, BOD calculations are automatic. When the BOD button is touched, a variety of meter screens will sequentially appear, allowing you to enter information essential to a BOD calculation. Among these are the sample description, ID number, and date. If a search of the memory does not match the current sample ID, the meter treats the sample’s DO reading as an initial reading.
Instruction Manual CyberScan DO5000 To Measure the BOD of the Seed Initial DO Measurement 1. Access the BOD mode from the main mode screen 2. Once the DO reading is stable, touch BOD. 3. Using the bottle ID numeric touch pad, key in a sample bottle ID number 4. Touch enter to accept the sample description. Measure-BOD std mg/L 8.21 Bottle ID STABLE 1 STND 123 BOD 2 3 4 5 6 7 8 9 Not Standardized setup Touch meas to measure sample or - 0 .
Instruction Manual CyberScan DO5000 5. If the sample is new, the BOD first reading screen appears. Touch SEED. Measure-BOD std mg/L 8.21 BOD first reading BOD Is this reading a SEED or a SAMPLE? STND Not Standardized setup Touch meas to measure sample or Touch std to access standardize mode August 12, 2002 SEED ID# auto cal ON 9.32am QUIT ATC 25.3°C Salinity 0.
Instruction Manual CyberScan DO5000 6. Key in a sample description using the alphanumeric touch pad, SEED, for example. 7. Touch enter to accept the sample description. enter Description SEED A B C 1 D E F 2 G H I 3 J K L 4 M N O 5 P Q R 6 S T U 7 V W X 8 Y Z - 9 BS / .
Instruction Manual CyberScan DO5000 8. Key in the volume of the SEED sample using the alphanumeric touch pad. Touch enter to accept the SEED volume. SEED volume 123 STABLE 1 2 3 4 5 6 7 8 9 - 0 . BS exit enter Not Standardized 9. Touch enter to accept the Seed dilution ratio. 10. Use the bottle volume touch pad to key in the mL volume of the BOD bottle. Bottle volume 123 STABLE 1 2 3 4 5 6 7 8 9 - 0 . BS exit enter Not Standardized 11.
Instruction Manual CyberScan DO5000 Final DO Measurement 1. Access the BOD mode from the main mode screen 2. Once the DO reading is stable, touch BOD. 3. Using the bottle ID numeric touch pad, key in a sample bottle ID number 4. Touch enter to accept the bottle ID number. If the meter’s barcode scanning feature is employed, simply scan the bottle to record its ID number.
Instruction Manual CyberScan DO5000 To Measure BOD of a Sample Initial DO measurement 1. Access the BOD mode from the main mode screen 2. Once the DO reading is stable, touch BOD 3. Using the bottle ID numeric touch pad, key in a sample bottle ID number 4. Touch enter to accept the bottle ID number. If the meter’s barcode scanning feature is employed, simply scan the bottle to record its ID number.
Instruction Manual CyberScan DO5000 5. If the sample is new, the BOD first reading screen appears. Touch SAMPLE. Measure-BOD std mg/L 8.21 BOD first reading BOD Is this reading a SEED or a SAMPLE? STND Not Standardized setup Touch meas to measure sample or Touch std to access standardize mode August 12, 2002 SEED ID# auto cal ON 9.32am QUIT SAMPLE ATC 25.3°C Salinity 0.0 ppt mmHg 732 6. Key in a sample description using the alphanumeric touch pad. 7.
Instruction Manual CyberScan DO5000 8. Use the Predilution Ratio touch pad to key in the ml ratio of sample to final volume. If there is no predilution, use the ratio 1:1, and key in the value 1. If the ratio is 1:100, key in 100. Predilution (1 to X) STABLE 1 123 2 3 4 5 6 7 8 9 - 0 . BS exit enter Not Standardized 9. Touch enter to accept the Seed Volume. 10. Use the Seed Volume touch pad to key in the mL volume of the sample. 11. Touch enter to accept the Seed Volume. 12.
Instruction Manual CyberScan DO5000 BOD Results 817 Seed ID: 1, SEEDA Predilution: Seed Volume: mg/L 1 3 ml Bottle Volume: Initial DO: Final DO: Delta DO: 300 ml 8.17 mg/L 0.00 mg/L 8.
Instruction Manual CyberScan DO5000 At this point, the initial DO and all of the above information about the sample will be stored in the data log. After an incubation period of at least one day, the DO of the sample must be measured again. Final DO Measurement 1. Access the BOD mode from the main mode screen. 2. Once the DO reading is stable, touch BOD. 3. Using the Manual ID numeric touch pad, key in the sample bottle ID number.
Instruction Manual CyberScan DO5000 6.3 Oxygen Uptake Rate (OUR) Measurement Measure - OUR std mg/L/hr 8.21 STABLE 0:00 STND start Not Standardized setup Touch start to begin measurement or Touch std to access standardize mode August 12, 2002 ID# autocal mg/L ON 1.28 9.32am ATC 25.3°C Salinity 0.0 ppt Altitude 732 Measure - OUR (RUNNING) mg/L/hr 8.
Instruction Manual CyberScan DO5000 1. Touch OUR on the main mode screen. 2. Touch setup to set parameters essential for the OUR calculation. These include dilution ratio, minimum run time, maximum run time, minimum DO required to end the test. 3. Touch start to initiate the OUR test. 4. Touch stop to terminate the test at any time. A timer on the OUR screen will display the time of the test. When the test concludes, the OUR results in mg/L/hr will be displayed.
Instruction Manual CyberScan DO5000 6.4 Specific Oxygen Uptake Rates (SOUR) Measure - SOUR std mg/hr/g 8.21 STABLE 0:00 STND start Not Standardized setup Touch start to begin measurement or Touch std to access standardize mode August 12, 2002 ID# autocal mg/L ON 1.28 9.32am ATC salinity mmHg 25.3°C 0.0 ppt 732 Measure - SOUR (RUNNING) mg/hr/g 1:35 STND 0.
Instruction Manual CyberScan DO5000 1. Touch SOUR on the main mode screen. 2. Touch setup to set parameters essential for the SOUR calculation. These include dilution ratio, minimum run time, maximum run time, minimum DO required to start the test, the minimum DO required to end the test and the total solids of the sample in g/L. 3. Touch start to initiate the SOUR test. 4. Touch stop to terminate the test at any time. A timer on the SOUR screen will display the time of the test.
Instruction Manual CyberScan DO5000 7 CLEANING The touch screen should be kept as clean as possible to preserve optical properties. Attempt to keep the screen free of dirt, dust fingerprints, etc. long term contact with abrasive materials will scratch the surface, and impair image quality. To clean, use a damp nonabrasive cloth towel and any commercially available window cleaner. The cleaning solution should be applied to the towel rather than the surface of the touch screen.
Instruction Manual CyberScan DO5000 9 DATA MANAGEMENT Your CyberScan DO 5000 meter is equipped with an RS-232 port for sending measurement data to a printer. The printer must be capable of receiving data within the following guidelines. Baud rates Data Bits Parity Stop Bits 110 to 38400 bps 7 or 8 even, odd or none 1 or 2 The meter and the external device must be configured to match with respect to these parameters. A shielded RS-232 cable must be used.
Instruction Manual CyberScan DO5000 To download data stored in memory to a printer, access the View Stored Data option in the setup screens. Access the desired data point by sorting the data via one of the three sorting options. Touch print on the screen displaying the data. The following is a list of factory default settings for the CyberScan DO 5000 meter. You can reset your meter to the factory default settings by accessing the Reset to Factory Default screen from the System Setup screen.
Instruction Manual CyberScan DO5000 10 DO FUNDAMENTALS Meter Operation The CyberScan DO 5000 meter uses the polarographic method of determining dissolved oxygen concentration. A voltage is maintained across two electrodes (a gold cathode and a silver anode) and an electrolyte separated from the sample solution by an oxygen permeable membrane. Oxygen diffuses form the sample across the membrane where it is reduced at the cathode.
Instruction Manual CyberScan DO5000 11 METER SPECIFICATIONS Feature Display Screen size Details DO 5000 PCD 5500 250 1000 640X480 back-lit dotmatrix LCD 114.3(W) X 152.4(H) mm Measurement display height 19.1 (H) mm Temp/etc. display height 6.
Instruction Manual CyberScan DO5000 Feature Details DO 5000 -2.000 to ±20.000 ا ~ 0.1/ 0.01/ 0.001 ~ Relative Accuracy ±0.002 ~ Calibration points 5 ~ KA,KS,AA,AS ~ pH mode Range Resolution Incremental methods ~ FET mV mode Range ±1800.0 ~ Resolution 0.1 ~ Accuracy ±0.1 ~ Ion mode Range 1 X 10-6 to 9.99 X 1010 Resolution 0.1/ 0.01/ 0.001 ~ ~ Relative accuracy ±0.17n% ~ Calibration points 5 ~ KA, KS, AA, AS ~ 0.1, 1.0, 10 ~ 0 to 300 mS/cm ~ 30 megohm.
Instruction Manual CyberScan DO5000 Feature Details DO 5000 General ~ Password Protection Input/ output Electrical requirements DIN/ RS232 2BNC/ 2 socket pin(reference, 2-pin conductivity)/ 2.5mm phono jack(ATC), 2DIN (ISFET, 4cell or DO) Input: 115VAC or 230 VAC Output: 12VDC, 500 mA (center negative) Line voltage tolerance Input impedance Meter size (Meter only) Meter size (Boxed) ± 10% >1012 ohms 165 (W) X 235 (L) X 89 (H) mm 280 (W) X 490 (L) X 160 (H) mm Meter weight (Meter only) 1.
Instruction Manual CyberScan DO5000 12 ACCESSORIES Consult your Authorized Distributors for these items and other range of specialized pH electrodes or Ion Selective Electrodes. 12.1 Replacement Meters and Accessories CODE NO.
Instruction Manual CyberScan DO5000 13 WARRANTY Eutech Instruments supplies the bench meters with a 3-year warranty and 6-month warranty for probes against manufacturing defects from the date of purchase. If repair or adjustment is necessary and has not been the result of abuse or misuse within the warranty period, please return, freight prepaid, and correction will be made without charge. Out of warranty items will be repaired on a charge basis.
NOTES
For more information on Eutech Instruments products, contact your nearest Eutech Instruments distributor or visit our website listed below: Manufactured by: Eutech Instruments Pte Ltd Blk 55, Ayer Rajah Crescent, #04-16/24 Singapore 139949 Tel: (65) 6778 6876 Fax: (65) 6773 0836 E-mail: marketing@eutechinst.com Web-site: www.eutechinst.