Instruction Manual PD 300
8.2 P1.0: Viewing previous pH calibration data
This mode lets you recall previous pH calibration data,
which helps you know when to re-calibrate your meter.
This is a “view only” mode.
From measurement mode:
1. Press the MODE key to select pH measurement
mode if necessary.
2. Press the SETUP key to enter Set up mode.
3. Press the or keys to scroll through
subgroups until you view parameter P1.0.
4. Press the ENTER key repeatedly to view previous
calibration data. See figure on right.
P1.1 = pH 1.68
P1.2 = pH 4.01
P1.3 = pH 7.00
P1.4 = pH 10.01
P1.5 = pH 12.45
5. When you have scrolled through all calibration
data, you will automatically return to the subgroup
menu. Press CAL/MEAS key to return to
measurement mode.
NOTE
If there is no previous calibration data at a particular
point, the primary display will show “ ----“.
Figure 21 – P1.0: View calibration