5. USER CALIBRATION
5.1. %Sat – One Point calibration (Auto)
User can do the 100%Sat calibration. The pressure that has been set
in the setup will be applied only during the calibration.
Accepted window for 100% sat= 50% - 200%Sat
1. Press std to enter to ‘Standardize’ screen. Upper display shows
the present measured value, while lower display shows the
pressure compensated 100% sat.
2. Hold probe in air.
3. Wait for the upper display value to stabilise. Press enter key to
confirm calibration. Meter blinks the cal values for few seconds
before exiting to measure screen.
press to accept
enter
Standardize
% Sat