8.8.2 P7.2: Selecting
°
C or
°
F
You can select between °C and °F units for temperature readings. Meter default is °C.
From measurement mode
1. Press the MODE key to select conductivity measurement mode.
2. Press the SETUP key to enter Setup mode.
3. Press the or key to scroll through sub-groups until you view
parameter P7.0.
4. Press the ENTER key two times to select parameter P7.2.
5. Press the or key to toggle between °C and °F.
6. Press the ENTER key to confirm selection and to proceed to step 4
of P7.3. Press the CAL/MEAS key to return to measurement mode.
NOTE
: You can switch between
°
C and
°
F in Program P3.4 (available
from pH mode). If you switch between
°
C and
°
F in conductivity mode,
the meter will also switch in pH mode.
8.8.3 P7.3: Selecting Automatic or Manual Temperature
Compensation
This feature lets you select between Automatic Temperature
Compensation (ATC) and Manual Temperature Compensation. Meter
default is ATC.
From measurement mode
1. Press the MODE key to select conductivity measurement mode.
2. Press SETUP key to enter Setup mode.
3. Press the or key to scroll through the sub-groups until you
view parameter P7.0.
4. Press the ENTER key three times to select parameter P7.3. The
upper display shows “ATC” and the lower display shows “P7.3”.
5. Press the ENTER key again. The upper display shows “ATC” and
the lower display shows “YES” or “NO”.
6. Press the or key to select the Automatic Temperature
Compensation on or off.
YES = ATC on; NO = ATC off
7. Press the ENTER key to confirm selection and proceed to step 3 of
P7.4. Press the CAL/MEAS key to return to measurement mode.
Figure 36: Change temperature
measurement unit
SETUP
C
p 7.2
°C
SETUP
F
p 7.2
°F
E
N
T
E
R
R
A
N
G
E
Figure 37: Selecting ATC
ON or OFF
SETUP
p 7.3
°C
ATC
E
N
T
E
R
R
A
N
G
E
SETUP
YES
°C
ATC
E
N
T
E
R
R
A
N
G
E
SETUP
°C