8.6 P5.0: Viewing previous conductivity calibration data
This mode lets you recall previous calibration data, which helps you know when to re-calibrate your
meter. This is a “view-only” mode.
This function applies for conductivity & TDS mode.
From conductivity or TDS mode
1. Press the MODE key to select conductivity or TDS measurement
mode.
2. Press the SETUP key to enter Setup mode.
3. Press the or key to scroll through sub-groups until you view
parameter P5.0.
4. Press ENTER key repeatedly to view previous calibration data.
5. Each calibration data corresponds to each measurement range.
• P5.1 = Range 1 (0.00 – 19.99 µS or 0.0 – 9.99 ppm)
• P5.2 = Range 2 (0.0 – 199.9 µS or 0.0 – 99.9 ppm)
• P5.3 = Range 3 (0 – 1999 µS or 0 – 999 ppm)
• P5.4 = Range 4 (0.00 – 19.99 mS or 0.00 – 9.99 ppt)
• P5.5 = Range 5 (0.0 – 199.9 mS or 0.0 – 99.9 ppt)
6. When you have scrolled through all calibration data, you will
automatically return to the sub-group menu. Press CAL/MEAS key
to return to measurement mode.
NOTE
: If there is no previous calibration data at a particular point, the
primary display will show “----“.