Preface This manual serves to explain the use of the Waterproof PC 10 hand-held meters. It functions as a step by step guide to help you operate the meter and as a handy reference guide. It is written to cover as many anticipated applications of the Waterproof PC 10 meters as possible. If there are doubts in the use of this meter, please do not hesitate to contact the nearest Authorised Distributor.
TABLE OF CONTENTS 1. INTRODUCTION ........................................................................................ 1 2. DISPLAY AND KEYPAD FUNCTIONS ................................................... 2 2.1. 2.2. 3. DISPLAY................................................................................................. 2 KEYPAD ................................................................................................. 3 PREPARATION ........................................................
Instruction Manual PC 10 1. INTR ODUCTION Thank you for selecting the Waterproof PC 10 portable meter. The meter is a microprocessorbased instrument that measures pH, conductivity, and temperature with just one probe. This meter has many user-friendly features – all of which are completely accessible through the water-resistant membrane keypad. Your meter includes 4 “AAA” batteries and a combination pH / conductivity / temperature probe with 3-meter submersible cable.
Instruction Manual PC 10 2. D ISPLAY AND K EYPAD FU NCTION S 2.1. Display The LCD has a primary and secondary display. • The primary display shows the measured pH or conductivity (µS or mS) reading. • The secondary display shows the temperature in °C. The display also shows error messages, keypad functions and program functions. See Figure 2.
Instruction Manual 2.2. PC 10 Keypad The large membrane keypad makes the instrument easy to use. Each button, when pressed, has a corresponding graphic indicator on the LCD. See Figure 3. Key Function ON/OFF Powers on and shuts off the meter. The meter powers on in the mode it was shut off in. For example, if you shut the meter off in Conductivity measurement mode, the meter will be in Conductivity measurement mode when you switch the meter on. HOLD Freezes the measured reading.
Instruction Manual PC 10 3. PR EPARATION 3.1 Inserting the Batteries The meter is packaged with 4 “AAA” alkaline batteries required for operation. To insert the batteries into the meter, follow the procedure outlined below. 1. Use a Philips screwdriver to remove the two screws holding the battery cover. See Figure 6 below. 2. Remove battery cover. 3. Insert batteries. Follow the diagram inside the cover for correct polarity. 4.
Instruction Manual 3.2 PC 10 Connecting the Probe The Waterproof PC 10 meter uses a special combination pH / Conductivity / Temperature probe. The probe cable has a notched 6-pin connector to attach the probe to the meter. NOTE: Do not substitute other probes or electrodes. For replacement probe, see the “Accessories” section, page 21-23. NOTE: Keep connector dry and clean. Do not touch connector with soiled hands. To connect the pH / Conductivity / Temperature probe: 1.
Instruction Manual PC 10 4. CA LIBRATION 4.1. Important Information on Meter Calibration When you recalibrate your meter, old pH and conductivity calibration data are replaced on a point by point basis. For example, if you previously calibrated your meter at pH 4.0, 7.0, and 10.0, and you recalibrate at pH 7.0, the meter retains the old calibration data at pH 4.0 and pH 10.0.
Instruction Manual 4.2. PC 10 Preparing the Meter for Calibration The pH/conductivity/temperature probe included with this meter is designed for use with this meter only. Do not substitute other types of probes or electrodes. For a replacement probe, see the “Accessories” section, page 21. Be sure to remove the protective cap of the pH probe before calibration or measurement. Always immerse the probe beyond the electrode band.
Instruction Manual 4.3. PC 10 Temperature Calibration The built-in temperature sensor included in the probe is factory calibrated. Calibrate your sensor only if you suspect temperature errors may have occurred over a long period of time or if you have a replacement probe. 1. Turn the meter on. Press MODE to select pH Measurement mode. 2. Press the CAL/MEAS Figure 7: Enter pH Calibration mode key to enter pH calibration mode. The CAL indicator will appear above the primary display.
Instruction Manual 6. PC 10 Press the ENTER key to confirm temperature calibration. The CON indicator will flash to confirm calibration. See Figure 13. NOTES To exit from Temperature Calibration mode without confirming calibration, DO NOT press ENTER in step 6. Press CAL/MEAS instead. Temperature calibration is restricted to ±5°C from Figure 10: CONFIRM temperature calibration the initial value displayed during calibration (shown in the secondary display).
Instruction Manual 4.4. PC 10 pH Calibration This instrument is capable of up to 3-point pH calibration to ensure accuracy across the entire pH range of the meter. You can perform 1, 2, or 3 point calibration with standard pH buffers of 4.01, 7.00 and 10.00. We recommend that you perform at least 2-point calibration using standard buffers that covers the expected sample range. You can also perform a 1-point calibration, but make sure that the buffer value is close to the sample value you are measuring.
Instruction Manual 6. PC 10 After the reading has stabilized and the READY indicator is displayed, press ENTER to confirm calibration. A confirming indicator (CON) flashes and disappears. The meter is now calibrated to the buffer indicated in the secondary display. See Figure 13. The secondary display automatically scrolls to the next buffer calibration option. If you are performing multi-point calibration, go to Figure 13: CONFIRM calibration point step 7.
Instruction Manual 4.5. PC 10 Conductivity Calibration The Important Information on Conductivity Calibration Your meter has four measuring ranges. You can calibrate one point in each of the measuring ranges (up to four points). If you are measuring values in more than one range, make sure to calibrate each of the ranges you are measuring. The following table lists the corresponding conductivity range.
Instruction Manual 4.6. 1. PC 10 Calibrating for Conductivity: Pour out two separate portions of your calibration standard and one of DI water into separate clean MEAS READY mS containers. 2. If necessary, press MODE key to select °C ATC Conductivity Mode. The µS and mS indicator will appear on the right side of the display. 3. Rinse your probe with DI water, then rinse the probe in one portion of the calibration standard.
Instruction Manual 6. PC 10 Press the ENTER key to confirm calibration. Upon confirmation, the CON indicator appears briefly. The meter automatically switches back into Measurement mode. The display now shows the calibrated, temperature compensated conductivity value. See Figure 21. 7. For calibration in other ranges (maximum: four ranges) repeat steps 1 through 6 with the appropriate calibration standards.
Instruction Manual PC 10 5 . M EA SUR EM EN T The READY indicator appears on the display when the readings stabilise. It will turn off if the readings MEAS pH start to fluctuate. NOTE: Be sure to remove the protective cap of the pH electrode before measurement. °C ATC To take readings: 1. Rinse the probe with DI water before use to remove any impurities adhering to the probe body. If the pH electrode has dehydrated, Figure 19: Measurement mode soak it for 30 minutes in tap water. 2.
Instruction Manual PC 10 6. HOLD FUNCTION This feature lets you freeze the value of the pH or conductivity reading for a delayed observation. HOLD can be used any time when in MEAS mode. 1. To hold a measurement, press the HOLD key while in measurement mode. “HOLD” will appear on the display. See Figure 25. 2. To release the held value, press HOLD again. Continue to take measurements.
Instruction Manual PC 10 7. PROB E CAR E AND MA INTENANC E Under typical operating conditions, the probe will need to be replaced every 6 to 12 months. In extreme applications, the probe may wear out sooner. Proper care and maintenance will help you receive the maximum probe life and ensure more accurate readings. Keep the probe clean. Before use, rinse the probe twice. Be sure to remove the protective cap of the pH probe before use and always immerse the probe beyond the electrode band.
Instruction Manual PC 10 8. TR OUBLESHOOTING Problem Cause Solution No display when turned on 1. Batteries not in place. 1. Check that batteries are in place and making good contact. 2. Batteries not in correct polarity (+ and -). 2. Reinsert batteries with correct polarity. 3. Weak batteries 3. Attach optional AC adapter (only for Standard PC 10 meter) or replace batteries. 1. Air bubbles in probe. 1. Tap probe to remove bubbles. 2. Dirty probe. 2. Clean the probe and recalibrate.
Instruction Manual PC 10 9 . E R R O R M E S S A G ES LCD Display Indicates Cause Solution Err annunciator Unrecognised input from keypad Wrong input in selected mode. Release key. Select valid operations depending on mode. CAL & Err annunciators blink. Calibration error. Wrong value input at calibration. Check your input value, clean probe. Dirty probe. See Calibration sections or Probe Maintenance section. Battery indicator blinks Low battery level.
Instruction Manual PC 10 1 0 . S P E C I F ICAT ION S Mode pH Temperature Conductivity Range 0.00 to 14.00 pH 0.0 to 100.0 °C 0 to 19.99 µS 0 to 199.9 µS 0 to 1999 µS 0 to 19.99 mS 0.01 pH 0.1 °C 0.01 µS 0.1 µS 1 µS 0.01 mS ± 0.01 pH ± 0.5 °C ±1% Full Scale +1 digit Up to three points (pH 4.01, 7.00, 10.01) with automatic buffer recognition Offset in 0.1 °C increments Up to four points (one point per range) Resolution Accuracy Calibration 1.0 Conductivity Cell constant (k) 2.
Instruction Manual PC 10 1 1 . A C C E S SO R I E S 11.1. Replacement Meter and Meter accessories Eutech Instruments Item Ordering Code Waterproof PC 10 pH / Conductivity / Temperature meter and electrode (EC-COMBI03M / 35630-50) EC-PCWP10/03K includes kit (EC-PCWP-KIT) Replacement pH / Conductivity / Temperature electrode.
Instruction Manual PC 10 Oakton Instruments Item Ordering Code Waterproof PC 10 pH / Conductivity / Temperature meter and electrode (EC-COMBI03M / 35630-50) 35630-02 Replacement pH / Conductivity / Temperature electrode.
Instruction Manual 11.2. PC 10 Calibration Solutions Note: pH buffer solutions have ±0.01 pH accuracy at 25 °C. Item Eutech Instruments Ordering Code Oakton Instruments Ordering Code pH 4.01 buffer solution, 480 mL bottle (1 pint) EC-BU-4BT 00654-00 pH 7.00 buffer solution, 480 mL bottle (1 pint) EC-BU-7BT 00654-04 pH 10.01 buffer solution, 480 mL bottle (1 pint) EC-BU-10BT 00654-08 pH 4.01 buffer pouches, 20 mL x 20 each EC-BU-4BS 35653-01 pH 7.
Instruction Manual PC 10 12. WARRANTY This meter is supplied with a three -year warranty, six-month warranty for probe against significant deviations in material and workmanship. If repair or adjustment is necessary and has not been the result of abuse or misuse within the designated period, please return – freight pre-paid – and correction will be made without charge. Eutech Instruments/ Oakton Instruments will determine if the product problem is due to deviations or customer misuse.
Instruction Manual PC 10 1 3 . R ET U R N O F I T EM S Authorisation must be obtained from our Customer Service Department or authorised distributor before returning items for any reason. A “Return Goods Authorisation” (RGA) form is available through our authorised distributor. Please include data regarding the reason the items are to be returned. For your protection, items must be carefully packed to prevent damage in shipment and insured against possible damage or loss.
For more information on Eutech Instruments/ Oakton Instruments’ products, contact your nearest distributor or visit our website listed below: Oakton Instruments P.O Box 5136, Vernon Hills, IL 60061, USA Tel: (1) 888-462-5866 Fax: (1) 847-247-2984 E-mail: info@4oakton.com Web-sites: www.4oakton.com www.oaktoninstruments.com Eutech Instruments Pte Ltd. Blk 55, Ayer Rajah Crescent, #04-16/24 Singapore 139949 Tel: (65) 6778 6876 Fax: (65) 6773 0836 E-mail: marketing@eutechinst.com Web-site: www.eutechinst.