8.7 P6.0: Memory Clear
Use this program to clear all memory values when you
need to store a new series of values. This is to avoid
confusion between the old values and the new ones.
From measurement mode:
1. Press the SETUP key to enter Set Up menu.
2. Press the MI/ or MR/ keys until you view the
parameter P6.0.
3. Press the HOLD/ENTER key to enter parameter
P6.0.
4. Press the MI/ or MR/ keys to toggle between
NO and YES.
NO retains current memory.
YES clears all memory.
5. Press the HOLD/ENTER key to confirm selection
and return to the subgroup menu.
6. Press CAL/MEAS key to return to measurement
mode.
Figure 40: P6.0-Clearance of
all stored memory