Instruction Manual CON 11 & CON 110 Handheld Conductivity/TDS/Temperature/RS232C Meter 68X361303 Rev. 1 - 12/03 Technology Made Easy ...
Preface This manual serves to explain the use of the CON 11/110 Conductivity/TDS handheld meters. It functions as a handy reference step by step guide to help you operate the meter. It is written to cover as many anticipated application of the CON 11/110 meters as possible. If there are doubts in the use of the meter, please do not hesitate to contact the nearest Authorised Distributor.
Table of Contents 1 INTRODUCTION ....................................................................................................................................... 1 2 DISPLAY AND KEYPAD FUNCTIONS.................................................................................................. 2 2.1 2.2 3 PREPARATION.......................................................................................................................................... 5 3.1 3.2 3.3 3.4 3.4.1 3.5 3.6 3.6.
8.3 P2.0: VIEWING CALIBRATION DATA .................................................................................................... 42 8.4 P3.0 VIEWING ELECTRODE DATA ....................................................................................................... 43 8.5 P4.0: AUTO OFF .................................................................................................................................. 44 8.6 P5.0: RESET TO FACTORY DEFAULTS .............................................
Instruction Manual 1 CON 11/110 INTRODUCTION Thank you for selecting the CON 11/110 Conductivity/TDS handheld meter. These meters are microprocessor-based instruments and are designed to be handy, capable of allowing one-hand operation. Each has a large customised LCD for clear and easy reading. It also has user-friendly features, all of which are accessible through the splash-proof membrane keypad.
Instruction Manual 2 CON 11/110 DISPLAY AND KEYPAD FUNCTIONS 2.1 Display The LCD has a primary and secondary display. • The primary display shows the measured Conductivity/TDS values. • The secondary display shows the measured temperature. The display also shows error messages, keypad functions and program functions. Primary Display 1 SETUP MEAS CAL 18 READY 17 16 HOLD 15 4 3 2 MEM 1.8.8.8 K= ON OFF ERR 14 13 12 1.8.8.
Instruction Manual 2.2 CON 11/110 Keypad A large membrane keypad with tactile feedback makes the instruments easy to use (Figure 2). Each button, when pressed, has a corresponding graphic indicator on the LCD. Some button has several functions depending on its mode of operation. KEY ON OFF FUNCTION • ON/OFF - Powers on and shuts off the meter. The meter will start in the measurement mode it was in when last switched off.
Instruction Manual CON 11/110 ON OFF CAL MEAS MI SETUP ON OFF HOLD ENTER CAL MEAS MR MI MODE SETUP RANGE HOLD ENTER MR MODE RANGE Keypad for CON 11 meter Keypad for CON 110 meter Figure 2: CON 11 and CON 110 keypads 4
Instruction Manual 3 CON 11/110 PREPARATION 3.1 Inserting the Batteries This meter is packaged with 4 “AAA” alkaline batteries required for its operation. To insert the batteries into the meter, follow the procedure outlined below. 1. Open the battery compartment by pressing down the catch of the battery cover. 2. Insert the batteries correctly according to the marked polarity of the battery compartment. 3. Close the battery cover, ensuring the catch is locked in position.
Instruction Manual 3.2 CON 11/110 Conductivity Electrode Information Your meter includes a conductivity electrode (Part No: ECCONSEN91W/ 35608-50) Ultem / Stainless Steel electrodes with an electrode constant of K = 1.0. This conductivity / TDS electrode features a built-in temperature sensor for Automatic Temperature Compensation (ATC). It has a specially designed housing that provides fast temperature response and reduces air bubble entrapment, which makes it easy to obtain accurate, stable readings.
Instruction Manual CON 11/110 Connecting the electrode to the meter 3.3 The CON 11 / 110 meter use the Conductivity / TDS electrode with a sturdy 6 pin connector which is easy to connect. 1. Insert the 6 pin female connector of the electrode to the 6 pins male connector on the meter. Rotate the locking ring clockwise until it locks. 2. To remove the electrode, simply rotate the connector’s locking ring counterclockwise and pull away gently for a complete removal.
Instruction Manual 3.4 CON 11/110 Attaching the Electrode Holder to the Meter The CON meter is designed to allow you to do a one hand operation for any measurement or calibration. For that purpose, one electrode holder is provided. They are designed for easy use and installation. Care must be taken to avoid use of excessive force in the process of attaching this component. 1. Locate the slot on the right side of the meter. 2. Gently slide the flange of the holder into the slot of the meter.
Instruction Manual 3.5 CON 11/110 Connecting the AC/DC Adapter Besides using the 4 “AAA” alkaline batteries as the power source, the CON 11/110 meter can also operate from the power mains using the AC/DC power adapter either at 120/220 VAC (sold separately) This is extremely useful if you have an AC power source available ( e.g.
Instruction Manual CON 11/110 3.6.1 RS232C Configuration The CON 110 meter has a 9 pin female RS232C connector with the following pin out. PIN NO. DESCRIPTION 1 - 2 Transmit Data 3 - 4 - 5 CTS (Clear to Send) 6 - 7 GND (Ground) 8 - 9 - A one to one connection can be made with a 9 pin RS232C port of the computer.
Instruction Manual 4 4.1 CON 11/110 CALIBRATION Important Information on Meter Calibration Your meter has five measuring ranges. You have an option of calibrating your meter in a single point calibration for all the five ranges or for better accuracy; you can calibrate one point in each of the measuring ranges (up to five points). This selection of single point calibration and multi point calibration can be set in the SETUP menu Program P9.2. See page 52 .
Instruction Manual 4.2 CON 11/110 Preparing the Meter for Calibration Before starting calibration, make sure you are in the correct measurement mode. For best results, select a standard value close to the sample value you are measuring. Alternatively use a calibration solution value that is approximately 2/3 the full-scale value of the measurement range you plan to use. For example, in the 0 to 2000 µS conductivity range, use a 1413 µS solution for calibration.
Instruction Manual 4.3 CON 11/110 Temperature Calibration Your electrode features a built-in temperature sensor. The temperature sensor is factory calibrated. Calibrate your sensor only if you suspect temperature errors may have occurred over a long period of time or if you have a replacement electrode. MEAS 1. 2. 3. 4. 5. 6. Switch the meter on. Press the CAL/MEAS key to enter conductivity or TDS calibration mode. The CAL indicator appears above the primary display.
Instruction Manual 4.4 CON 11/110 Selection of Automatic or Manual Calibration This meter is capable of performing either automatic conductivity calibration or manual conductivity / TDS calibration. In the automatic calibration mode, the meter automatically detects and verifies the appropriate known calibration standards solutions before accepting these particular calibration standards as one of its calibration values in a specific measurement range.
Instruction Manual 4.5 CON 11/110 Automatic Calibration (For Conductivity Calibration Only) In the Automatic Calibration mode, the meter is capable of accepting either singlepoint or up to 4 points for multi-point calibration with maximum of 1 point per specific measurement range. Set up the automatic calibration mode in the SETUP menu Program P9.1 by setting to “YES”. See page 51 for the set up procedure.
Instruction Manual CON 11/110 IMPORTANT NOTES: 1. Meter allows a tolerance range of ±40% of its calibration standard. If the calibration standard is detected out of the tolerance range during the calibration mode, the meter will prompt by the toggling of the secondary display continuously between various calibration standard values. An error message “ERR” will be displayed for 3 seconds if you attempt to calibrate with a solution whose value is outside the tolerance range.
Instruction Manual 4.6 CON 11/110 Manual Calibration (For Conductivity & TDS Calibration) In Manual Calibration mode, you can use customised Conductivity / TDS calibration standards (specific to your own application) and calibrate the meter. The following example shows the calibration sequence to 12.00 mS conductivity calibration standard. Set up the manual calibration mode in the SETUP menu Program P9.1 by setting to “NO”. See page 51.
Instruction Manual 4.7 CON 11/110 TDS Calibration 4.7.1 Calibrating for TDS Using Conductivity Standards & adjusting TDS factor It is important to note that the increase in concentration of salts dissolved in solution increases the conductivity of that solution. This relationship varies from salt to salt and is roughly linear over a given range for a given salt. The TDS conversion factor is the number used by the meter to convert from conductivity to TDS.
Instruction Manual CON 11/110 4.7.3 Calibrating for TDS using TDS standards After setting the correct TDS Factor, you can commence calibration in the TDS mode. From the measurement mode, 1. If necessary, press the MODE key to select the TDS mode. 2. For the rest of the calibration process, repeat the similar step 2 to 6 as in the section of Manual Calibration on page 17, this time using the TDS calibration standards. NOTE: You can offset the TDS reading up to ± 40% from the default setting.
Instruction Manual 5 CON 11/110 MEASUREMENT This meter is capable of taking measurements with automatic (ATC) or manual (MTC) temperature compensation. Factory default is ATC on. 5.1 Automatic Temperature Compensation For automatic temperature compensation (ATC), simply attach the conductivity/TDS electrode to the meter. Ensure that the ATC mode has been selected in the SETUP menu. The ATC indicator will light on the LCD.
Instruction Manual 5.2 CON 11/110 Manual Temperature Compensation To select the Manual Temperature Compensation (MTC) option, set up the Manual Temperature Compensation mode in the SETUP menu Program P 1.3 by setting to ‘NO”. 1. Press SETUP key to enter Set Up mode. The display shows “COF” at “P 1.0”. 2. Press the HOLD/ENTER key to scroll through subgroups until you view “ATC” at “P1.3” with small annunciators of “°C” and “ATC” or “°F” and “ATC”. 3. Press the HOLD/ENTER key again.
Instruction Manual CON 11/110 5.2.1 Setting the manual temperature compensation value To use manual temperature compensation (MTC), you need to enter the temperature value of your sample into the meter. This is the temperature value at which the measured readings will be manually temperature compensated. You can select any temperature between -10.0 to 110.0 °C (14.0 to 230 °F). Default value is 25 °C. 1. Switch the meter on. If necessary, set up the MTC by selecting ATC off as described in page 21.
Instruction Manual 5.3 CON 11/110 Taking Measurements 1. Rinse the electrode with de-ionised or distilled water before use to remove any impurities adhering to the electrode body. Shake or air dry. To avoid contamination or dilution of your sample, rinse electrode with a small volume of your sample liquid. 2. Press ON to switch on meter. 3. Dip the electrode into the sample. When dipping the electrode into the sample, take care to ensure that the liquid level is above its upper steel band.
Instruction Manual 5.4 CON 11/110 Using Manual Ranging Function Your meter automatically selects the range in which your readings appear. The manual ranging function lets you select the specific range you want to work in: Meter Range Symbol Conductivity Range TDS Range r1 0 to 19.99 µS/cm, 0.00 to 9.99 ppm r2 20.0 to 199.9µS/cm 10.0 to 99.9 ppm r3 200 to 1999 µS/cm; 100 to 999 ppm r4 2.00 to 19.99 mS/cm, 1.00 to 9.99 ppt r5 20.0 to 199.9 mS/cm 10.0 to 99.9 ppt 5.4.
Instruction Manual 5.5 CON 11/110 HOLD Function This feature lets you freeze the display for a delayed observation. HOLD/ENTER can be used any time in MEAS mode. 1. To hold a measurement, press the HOLD/ENTER key while in measurement mode. “HOLD” will appear on the display. 2. To release the held value, press the HOLD/ENTER again. Continue to take measurements. Figure 21: HOLD freezes measurement NOTE: • If the meter is shut off either automatically or manually, the HOLD value will be lost.
Instruction Manual 6 CON 11/110 MEMORY FUNCTION 6.1 Memory Input CON 11/110 meter stores up to 50/100 set of data respectively. Data sets can include both conductivity and TDS measurement values and its temperature. 1. 2. MEAS 823 READY 21.3 During any measurement function (MEAS), press MI/▲ key to input any data into the memory. MEM, “StO” and memory number will appear for a moment after which the meter then returns to measurement mode.
Instruction Manual 6.2 CON 11/110 Memory Recall This function recalls the previous readings stored in the memory. You can access memory recall function from the measurement mode only. Memory recall is in "Last In First Out" order. 1. Press the MR/▼ key once to retrieve the last reading stored. The memory location screen – MEM, “LOC” and the memory number – will be on the display. 2. Press the HOLD/ENTER key to recall the reading stored under that memory number. 3.
Instruction Manual 7 CON 11/110 PRINT FUNCTION (FOR CON 110 ONLY) The CON 110 meter is equipped with an RS232 communication port and a PRINT ( ) key that facilitate the printing of data from the meter to a printer or computer. 7.1 Using CON 110 Meter with Printer or Computer 1. To use the CON 110 directly with a printer, the printer should have either a 9 pin or a 25 pin RS 232C serial port. 2.
Instruction Manual 7.2 CON 11/110 Sending Data To Computer or Printer When the PRINT ( ) key is pressed, there are two options: 1. dAt / CUr: Prints current data on timed interval basis. The function enables the meter to continuously print the current data automatically based on the pre-set timing from 5 to 30 seconds. 2. dAt / StO: Prints stored data from memory. This function lets you print all the memorised data from memory previously stored in the meter.
Instruction Manual CON 11/110 7.2.1 Printing Current Data Manually MEAS To manually print any data that is currently being measured: 1. READY 1413 From measurement mode, press HOLD/ENTER key to freeze reading. ) key. 2. Press PRINT ( 3. The “Prn – CUr” annunciator will appear and flash for about 0.5 seconds. The data is transmitted to the printer or computer. 22.3 µS °C ATC HOLD ENTER 4. The meter returns to measurement mode. 5. Repeat steps 1 to 3 for next measurement.
Instruction Manual CON 11/110 "MEM" annunicator will keep flashing during measurement mode & data will be printed on pre-set timed interval. To stop, press PRINT key From measurement mode MEAS MEM µS 1413 READY 21.0 HOLD ENTER °C ATC Print on set timed interval sec 5 MR Flash for 0.5 seconds MI MEAS sec 10 HOLD ENTER READY Set timed interval for printing, eg, every 10 seconds Figure 27: Print data on timed interval 31 MEM 1413 21.
Instruction Manual CON 11/110 7.2.3 Print Data from Stored Memory The CON 110 meter can print data that is stored in the memory. You can print out all stored data or selected data based on its memory location. To Print All Stored Data 1. From measurement mode, press PRINT ( ) key. 2. Use MI/▲ or MR/▼ key, select “dAt – StO” option and press HOLD/ENTER key. 3. At the “SEC – 0” option, use MI/▲ or MR/▼ key to select number of seconds to print each data stored data.
Instruction Manual CON 11/110 To print Selected Stored Data 1. From measurement mode, press MR/▼ key. 2. A “MEM” indicator appeared and the last memory input location is displayed. 3. Press MI/▲ or MR/▼ key to select a specific memory location, and press HOLD/ENTER key. 4. When ready, press PRINT ( 5. The screen flashes “Prn/CUr” for about 0.5 seconds and switches back to the stored data of the specific memory location. 6. Press HOLD/ENTER key.
Instruction Manual 8 CON 11/110 ADVANCED SETUP FUNCTION The advanced setup mode lets you customised your meter’s preferences and defaults. The CON 11/110 meter features different sub-groups that organise setup parameters. The sub-groups are: 1. P1.0: Unit Configuration (COF) 2. P2.0: Viewing previous calibration data (CAL) 3. P3.0: Viewing electrode data (ELE) 4. P4.0: Auto Power OFF (AtO) 5. P5.0: Reset to factory defaults (rSt) 6. P6.0: Memory Clear (CLr) 7. P7.
Instruction Manual CON 11/110 SETUP Asc SETUP MR P9.0 SETUP COF MI P1.0 MI SETUP CEL MR P8.0 SETUP MI CAL P2.0 MI MR SETUP MR P7.0 °C °F SETUP MI ELE P3.0 MI MR MEM SETUP MR P6.0 SETUP MI A o P4.0 MR SETUP MR MI P5.
Instruction Manual 8.1 CON 11/110 Advanced SETUP Mode Overview 1. Press the SETUP key to enter Set Up menu. 2. Press the MI/▲ or MR/▼ keys to scroll through sub groups. 3. Press HOLD/ENTER key to enter a particular parameter. P1.0: Unit Configuration P1.1 Selection of READY and *Auto HOLD function (*CON 110 Only). P1.2 Selection of °C or *°F (*CON 110 Only) P1.3 Selection of Automatic or Manual Temperature Compensation P1.
Instruction Manual SETUP CEL CON 11/110 P8.0: Selection of cell constant P8.0 Selection of cell constant: K = 1.0, 10, or 0.1 P8.0 SETUP Asc P9.0 P9.0: Automatic and Single Point Calibration P9.1 Selection of Automatic or Manual Calibration (This mode only appears when SETUP Menu is chosen while in the Conductivity measurement mode P9.
Instruction Manual 8.2 CON 11/110 P1.0: Unit Configuration (COF) 8.2.1 P1.1: Selection of READY and Auto HOLD function Program P1.1 gives you an option to on or off the “READY” display which indicates the measurement stability during a measuring process. Program P1.1 also gives an option to activate the Auto HOLD function (Only for CON 110). Activating the auto HOLD function lets the meter to “hold” your measurement when it is stable for more than 5 seconds.
Instruction Manual CON 11/110 8.2.2 P1.2: Selection of °C or °F Available in CON 110 only You can select between °C and °F units for temperature readings. Meter default is °C. HOLD ENTER From measurement mode 1. Press SETUP key to enter Set Up menu. 2. Press the HOLD/ENTER keys until you view parameter P1.2. 3. Press the MI/▲ or MR/▼ key to select between °C and °F. 4. Press the HOLD/ENTER key to confirm selection. 5. Press the CAL/MEAS key to return to measurement mode. SETUP C p 1.
Instruction Manual CON 11/110 8.2.3 P1.3: Selection of Automatic or Manual Temperature Compensation This feature lets you select between Automatic Temperature Compensation (ATC) and Manual Temperature Compensation (MTC). Meter default is ATC. From measurement mode 1 Press SETUP key to enter Set Up menu. 2 Press the HOLD/ENTER parameter P1.3. 3 4 keys SETUP until you view Press the HOLD/ENTER key again. The upper display shows “ATC” and the lower display shows “YES” or “NO”.
Instruction Manual CON 11/110 8.2.4 P1.4: Selection of TDS Conversion Factor You can set the TDS conversion factor between 0.40 and 1.00; meter default is 0.50. SETUP NOTE: Your meter has to be in the TDS measurement mode before you can make the selection of the TDS Conversion Factor in the SETUP menu. Only from TDS measurement mode 1. Press SETUP key to enter Set Up menu. 2. Press the HOLD/ENTER keys until you view parameter P1.4. 3. 4. Press the HOLD/ENTER key again.
Instruction Manual 8.3 CON 11/110 P2.0: Viewing calibration data This mode lets you recall the previous calibration data on each of the five measuring ranges. This is a “view only” mode. From measurement mode: SETUP cal 1. Press the SETUP key to enter Set Up menu. 2. Press the MI/▲ or MR/▼ key until you view parameter P2.0. 3. Press the HOLD/ENTER key repeatedly to view the previous calibration data on each of the five measuring ranges. 4.
Instruction Manual 8.4 CON 11/110 P3.0 Viewing Electrode Data Program 3.0 has five “view only” options that let you check the electrode’s parameters for diagnostic purposes. These options show you the effective cell constant for each range. The cell constant is adjusted SETUP according to your calibration. ele From measurement mode 1. Press the SETUP key to enter Set Up menu. 2. Press the MI/▲ or MR/▼ keys until you view parameter P3.0. 3.
Instruction Manual 8.5 CON 11/110 P4.0: Auto Off In Program P4.0 you can activate the Auto Off option to automatically switch off the instrument 20 minutes after the last key selection. This feature is useful for conserving battery power. Under default conditions, the SETUP instrument has this option activated. 1. Press the SETUP key to enter Set Up mode. 2. Press the MI/▲ or MR/▼ keys until you view parameter P4.0. 3. Press the HOLD/ENTER key.
Instruction Manual 8.6 CON 11/110 P5.0: Reset to factory defaults Program 5.0 reset functions lets you have an option to only reset all the meter’s calibrations and its data that have been stored or reset all calibrations, its stored data, memory, and any other setup functions that you might have changed back to factory default settings. See Factory Default Settings at page 69 and 70. 8.6.1 P5.
Instruction Manual CON 11/110 8.6.2 P5.2: User Reset This reset function clears all the calibrations, memory and all other setup parameters that you have changed back to factory defaults condition. SETUP cal From measurement mode 1. Repeat step 1 to 4 to enter the calibration reset selection mode. 2. Press the MI/▲ or MR/▼ keys to select NO. 3. Press the HOLD/ENTER key to enter parameter P5.2. The upper display will show “USr”. 4.
Instruction Manual 8.7 CON 11/110 P6.0: Memory Clear Use this program to clear all memory values when you need to store a new series of values. This is to avoid confusion between the old values and the new ones. SETUP From measurement mode: 1. Press the SETUP key to enter Set Up menu. 2. Press the MI/▲ or MR/▼ keys until you view the parameter P6.0. 3. Press the HOLD/ENTER key to enter parameter P6.0. 4. Press the MI/▲ or MR/▼ keys to toggle between NO and YES. 5. 6.
Instruction Manual 8.8 CON 11/110 P7.0: Temperature 8.8.1 P7.1: Setting the temperature coefficient The temperature coefficient is the amount of change in conductivity per degree of temperature; it is expressed in percent per °C.. Entering the exact temperature coefficient of your solution lets you accurately compensate SETUP temperature for almost any solution. You can adjust 0.0 to 10.0 % per °C. Meter default is 2.1% per °C. By setting the temperature coefficient to 0.
Instruction Manual CON 11/110 8.8.2 P7.2: Setting the normalisation temperature Your meter will normalise its conductivity measurements to a standard temperature that you can select. You can adjust the normalisation temperature either at 20.0°C or 25.0°C. Meter default is 25.0°C. SETUP p 7.2 From measurement mode 1. Press SETUP key to enter Set Up menu. 2. Press the MI/▲ or MR/▼ keys until you view parameter P7.0. 3. 4. HOLD ENTER SETUP 25.
Instruction Manual 8.9 CON 11/110 P8.0: Selection of cell constant Your meter lets you select a cell constant of K = 1.0, 10, or 0.1. Use a cell of K = 1.0 for midrange measurements Use a cell of K = 10 for high range measurements (above 20 mS or 10 ppt). Use a cell of K = 0.1 for low range measurements (below 20 µS or 10 ppm). SETUP cel The cell included with your meter has a cell constant of K = 1.0. p 8.0 From measurement mode 1. Press SETUP key to enter Set Up menu. 2.
Instruction Manual CON 11/110 8.10 P9.0: Automatic and Single point Calibration You may select an automatic or manual calibration depending on the calibration standards that you have. Program 9.0 lets you have the option to do the above as well as the option of having a single or a multi point calibration. 8.10.1 P9.
Instruction Manual CON 11/110 8.10.2 P9.2: Selection of Single or Multi Point Calibration The single point calibration lets you have a single calibration factor in all the five measuring ranges by just calibrating one point in either one of the five measuring ranges.
Instruction Manual 9 CON 11/110 CYBERCOMM PORTABLE DAS (FOR CON 110 ONLY) The DAS software is designed for pH 110 and CON 110 meters to allow you a convenient means of capturing data for future analysis using other software program such as LOTUS 123, EXCEL or DBASE in Windows©. Often one finds it cumbersome to record and transfer data from one media to another before the required processing can be done. With the DAS software, this redundant processing can be eliminated or reduced. 9.
Instruction Manual CON 11/110 Figure 47: Click on 'Browse' button and locate CD-ROM drive Figure 48 : Locate the CyberComm Portable Setup program in the CD-ROM under "Eutech DAS" directory.
Instruction Manual CON 11/110 Figure 49: Select ‘Portable Meters’ directory to go to CyberComm Portable. Figure 50 : Open ‘CyberComm Portable’ to enter to the Setup program.
Instruction Manual CON 11/110 Figure 51: Select "SETUP.EXE" program and click the OPEN button Figure 52: InstallShield Wizard dialog box appears.
Instruction Manual CON 11/110 Figure 54: Key in your name and company name and click NEXT button Figure 55: To select another Destination Directory to install the program, click on BROWSE button. Otherwise, click NEXT button.
Instruction Manual CON 11/110 Figure 56: Creating a new program folder. Click on NEXT button. Figure 57: Click on NEXT button.
Instruction Manual CON 11/110 Figure 58: The CyberComm Portable DAS program is fully installed. Click on FINISH button to end installation. 9.3 Running CyberComm Portable Before running the DAS program, please ensure that the RS232 cable is connected between the computer’s serial port and the meter’s port. See section 3.6- Connecting the RS232C Cable (Only for CON 110) on page 9.
Instruction Manual CON 11/110 Figure 60: The opening screen will appear as above 9.3.1 Buttons & Check-Box 1. Enable Connection - Click this button to enable communication between meter and computer. 2. Clear Readings - To clear all data and start all over again. 3. Save Readings - To save all data displayed in either *.dat or *.txt format. 4. Time Stamp - To include Time and Date stamp when collecting the data. Time and date information comes from the computer.
Instruction Manual CON 11/110 Figure 61: Under File Menu setting, you can change various parameters. Under ABOUT menu, details of Eutech Instruments' contact information, email address and updates are shown. 9.3.2 Menu 1. Communication Settings - To set communication port number, baud rate speed, parity and stop bits protocol. 2. Open - To open previously saved data file. 3. Save - To save current data captured. 4. Save As - To save current data set in another format such as *.dat or *.txt. 5.
Instruction Manual CON 11/110 Figure 62: Communication Settings for computer's Com port. It must match with COM port settings on CyberScan CON 110 meters. Please refer to Section7.1 “Using CON 110 Meter with Printer or Computer”on page 28 for the settings. 9.3.3 Communication Settings 1. Connect Using - For selecting communication port. (Select ‘1’) 2. Baud Rate – For selecting baud rate. (Select ‘9600’ bps (bits per second)). 3. Parity – For selecting parity. (Select ‘None’) 4.
Instruction Manual 9.4 CON 11/110 Capturing And Printing Data Into Computer Using Data Acquisition After matching the Communication Settings between your computer using Data Acquisition program and the CON 110 meter, you can now capture data into your computer for analysis and storage purposes. 1. Ensure the 1-meter RS232 communication cable (Refer to Accessories Section) is connected between the computer and the CON 110 meter’s Com port. Refer to section 3.
Instruction Manual 9.5 CON 11/110 Trouble-shooting Guide Problem: Unable to PRINT When press PRINT key on CON 110 meter, nothing is printed. POSSIBLE CAUSES SOLUTIONS You have not "ENABLE CONNECTION" in the Data Acquisition program. Click on "ENABLE CONNECTION" in the Data Acquisition program The "Communication Settings" in the Data Acquisition program is different from meter's setup.
Instruction Manual 10 CON 11/110 ELECTRODE CARE AND MAINTENANCE Keep the conductivity electrode clean. Rinse the electrode twice, and gently swirl it while you take readings. For best accuracy, soak a dry electrode for at least 5 to 10 minutes or longer before calibration. Rinse the electrode with de ionised or tap water before storing. Never scratch the bands with a hard substance. Do not strike the electrode against any hard surface. Do not make continuous contact with your solutions.
Instruction Manual 11 CON 11/110 ERROR MESSAGES The following table provides a guideline to enable diagnosis of possible problems indicated by the messages generated by the meter. The table also provides possible solutions to the problems encountered. ERROR MESSAGE INDICATES POSSIBLE CAUSE CORRECTIVE ACTION ERR. Annunciator Wrong keypad input Wrong input in selected mode. Release key. Select valid operations depending on mode.
Instruction Manual 12 CON 11/110 TROUBLE SHOOTING Problem Cause Solution No display when ‘ON’ key is pressed a) Batteries not in place b) Batteries not in correct polarity (+ and – position) c) Weak batteries d) No AC power to adapter (if this optional accessory is used) a) Check that batteries are in place and making good contact b) Re-insert batteries with correct polarity c) Replace batteries d) Check power to AC adapter Unstable readings a) Air bubbles in electrode.
Instruction Manual 13 CON 11/110 LIST OF ACCESSORIES 13.1 Replacement Meter and Meter accessories Eutech Instruments Ordering Code Oakton Instruments Ordering Code EC-CON11/03 35607-40 CyberScan CON 11 Hand-held conductivity /TDS meter complete with conductivity electrode of K=1.0 (ECCONSEN91W) in rugged hard-plastic carrying case kit (ECCONWPKIT/ 35632-98).
Instruction Manual 14 CON 11/110 FACTORY DEFAULT SETTINGS 14.1 CON 11 Factory Default Setting PROGRAM FUNCTION PARAMETER OPTING KEYS PARAMETER OPTIONS DEFAULT SETTING P 1.0 Unit Configuration Parameter Title Only Parameter Title Only P 1.1 Selection of READY Indicator MI/▲; MR/▼ ON ; OFF, ON P 1.2 °C indicator Viewing only Viewing only °C P 1.3 Selection of Automatic or Manual Temperature Compensation MI/▲; MR/▼ YES, nO YES P 1.
Instruction Manual CON 11/110 14.2 CON 110 Factory Default Setting PROGRAM FUNCTION PARAMETER OPTING KEYS PARAMETER OPTIONS DEFAULT SETTING P 1.0 Unit Configuration Parameter Title Only Parameter Title Only P 1.1 Selection of READY Indicator and Auto HOLD Function MI/▲; MR/▼ ON ; OFF, HOLD ON P 1.2 Selection of °C and °F indicator MI/▲; MR/▼ °C, °F °C P 1.3 Selection of Automatic or Manual Temperature Compensation MI/▲; MR/▼ YES, nO YES P 1.
Instruction Manual 15 CON 11/110 SPECIFICATIONS SPECIFICATIONS CON 11 CON 110 Conductivity Range 0 to 19.99 µS/cm, 20.0 to 199.9µS/cm, 200 to 1999 µS/cm; 2.00 to 19.99 mS/cm, 20.0 to 199.9 mS/cm 0 to 19.99 µS/cm, 20.0 to 199.9µS/cm, 200 to 1999 µS/cm; 2.00 to 19.99 mS/cm, 20.0 to 199.9 mS/cm TDS Range 0.00 to 9.99 ppm 10.0 to 99.9 ppm 100 to 999 ppm; 1.00 to 9.99 ppt 10.0 to 99.9 ppt 0.00 to 9.99 ppm 10.0 to 99.9 ppm 100 to 999 ppm; 1.00 to 9.99 ppt 10.0 to 99.9 ppt Resolution 0.
Instruction Manual CON 11/110 Auto HOLD function Not available Selectable Averaging/Stability (READY) Selectable Selectable Inputs 6-pin connector, power jack 6-pin connector, power jack Output No RS232C Ambient Temperature Operating Range 0 to 50 °C 0 to 50 °C Display Custom Dual LCD Custom Dual LCD Power Requirements 4 ‘AAA’ Batteries 4 ‘AAA’ Batteries AC/DC Adapter ratings I/P Voltage range 9 to 15V, Current rating 200 mA I/P Voltage range 9 to 15V, Current rating 200 mA Battery
Instruction Manual 16 CON 11/110 CALIBRATION TIPS You only need one calibration for measurement throughout the entire range of the meter. If a range was not calibrated, the meter automatically detects the closest range calibrated and uses that calibration information. However, only the ranges that were calibrated have maximum accuracy.
Instruction Manual 18 CON 11/110 CALCULATING TEMPERATURE COEFFICIENTS To determine the temperature coefficient of your sample solution use this formula: Where: tc = Temperature coefficient 25 = 25 °C CT1 = Conductivity at Temp 1 CT2 = Conductivity at Temp 2 T1 = Temp 1 T2 = Temp 2 NOTE: A controlled temperature water bath is ideal for this procedure 1.
Instruction Manual 19 CON 11/110 WARRANTY This meter is supplied with a three-year warranty, six-month warranty for electrode against significant deviations in material and workmanship. If repair or adjustment is necessary and has not been the result of abuse or misuse within the designated period, please return – freight pre-paid – and correction will be made without charge. Eutech Instruments/ Oakton Instruments will determine if the product problem is due to deviations or customer misuse.
For more information on Eutech Instruments/ Oakton Instruments’ products, contact your nearest distributor or visit our website listed below: Oakton Instruments P.O Box 5136, Vernon Hills, IL 60061, USA Tel: (1) 888-462-5866 Fax: (1) 847-247-2984 E-mail: info@4oakton.com Web-sites: www.4oakton.com www.oaktoninstruments.com Eutech Instruments Pte Ltd. Distributed by: Blk 55, Ayer Rajah Crescent, #04-16/24 Singapore 139949 Tel: (65) 6778 6876 Fax: (65) 6773 0836 E-mail: marketing@eutechinst.