13 SPECIFICATIONS OF CYBERSCAN CON 10/100/200
Models CON 10 CON 100 CON 200
Conductivity range
0 to 19.99, 199.9,
1999
µ
S/cm;
19.99 mS/cm
0 to 19.99, 199.9, 1999
µ
S/cm;
19.99, 199.9 mS/cm
Accuracy
±
1% Full Scale
Resolution 0.05% of Full Scale
TDS range
0 to 9.99, 99.9,
999 ppm; 9.99 ppt
0 to 9.99, 99.9,
999 ppm;
9.99, 99.9, 199 ppt
Resolution
0.01, 0.1, 1 ppm;
0.01 ppt
0.01, 0.1, 1 ppm/ppt
Temperature range 0.0 to 100.0
°
C
Accuracy
±
0.5
°
C
Resolution 0.1
°
C
TDS factor 0.5
0.40 to 1.00
(selectable)
No. of calibration
points
4, maximum 1 per
range
5, maximum 1 per range
Auto-ranging
capability
Yes
Averaging/stability Yes
Auto-off Yes
Memory 16 sets
User customization Yes
HOLD function Yes
Cell constant 1.0 0.1, 1.0 & 10.0
Temperature
coefficient
2 % per
°
C 0.0 to 10 % per
°
C (selectable)
Temperature
compensation
Auto/manual (from 0 to 80
°
C)
Display Custom LCD
Inputs BNC, phono, power jack
Printing option Yes
Outputs RS 232 port/printer
Power 4 AAA batteries; 9 V DC adapter
Battery life > 200 hours