User Guide
31
8.3.3 P2.2: Calibration Reset
P2.2 provides the option to reset all user selectable
functions to the default settings. If you toggle the
annunciator to ON using the MI/
!
or MR/
"
key, and press
ENTER/RANGE key, all data will be reset.
RESET clears
all calibration data in memory (of both conductivity
and TDS)
. You need to power ON the instrument before
proceeding with any other functions.
Under default condition, RESET option is OFF.
NOTE: Data cannot be selectively deleted.
ON Reset all calibration points
OFF Reset off
8.3.4 P2.3: Temperature Normalization
The ATC (Automatic Temperature Compensation)
function automatically compensates for the variation
caused by temperature difference during measurement
and at the standard temperature (20 or 25
°
C) and gives
the corrected read-out of the equivalent conductivity of
the solution. Hence, the conductivity/TDS can be
normalized to display the equivalent conductivity/TDS at
either 20 or 25
°
C.
Normalization temperature can be selected between 20 &
25
°
C using the MI/
!
or MR/
"
key. The primary display
shows normalization temperature. Pressing
ENTER/RANGE key registers the selected temperature at
which the conductivity/TDS value is normalized. Default is
at 25
°
C.
Fi
g
ure 28: The RESET o
p
tion
Figure 29: The Temperature
Normalization option