Instruction manual
Table Of Contents
- 1. INTRODUCTION
- 2. GETTING STARTED
- 3. CONDUCTIVITY ELECTRODE
- 4. PH AND MV CALIBRATION
- 5. CONDUCTIVITY AND TDS CALIBRATION
- 6. CONDUCTIVITY AND TDS MEASUREMENT
- 7. HOLD FUNCTION
- 8. STORING AND RECALLING DATA
- 9. SETUP FUNCTIONS
- 9.1 1.0 CAL (Calibration)
- 9.2 2.0 ELE (Electrode Information)
- 9.3 3.0 ConF (Configuration)
- 9.4 3.1 rdY (Ready / Stability Indicator)
- 9.5 3.2 ºC ºF (Celsius or Fahrenheit)
- 9.6 3.3 buFF (pH Buffers & Calibration Points)—pH only
- 9.7 3.3 AtC (Auto Temp Compensation)—Con & TDS only
- 9.8 3.4 tdS (TDS factor)—Con & TDS only
- 9.9 3.5 t.CO (Temperature Coefficient)—Con & TDS only
- 9.10 3.6 t.nr (Normalization Temperature in ºC)—Con & TDS only
- 9.11 3.7 ACAL (Auto Conductivity Calibration)—Con & TDS only
- 9.12 3.8 SPC (Single Point Calibration)—Con & TDS only
- 9.13 3.8 CELL (Nominal Cell Constant)—Con & TDS only
- 9.14 4.0 rSt (Reset)
- 9.15 5.0 CLr (Clear Memory)
- 10. CALCULATING TDS CONVERSION FACTOR
- 11. CALCULATING TEMPERATURE COEFFICIENTS
- 12. REPLACEMENTS AND ACCESSORIES
- 13. TROUBLESHOOTING GUIDE
- 14. SPECIFICATIONS
- 15. WARRANTY
- 16. RETURN OF ITEMS
- info@4oakton.comwww.4oakton.com

7. HOLD FUNCTION
For prolonged observation of a reading, press during
measurement mode to freeze the display. The “HOLD” indicator will
display when the reading is held. To release the held value and
resume measurement, press
again or insert the held value into
memory by pressing
.
8. STORING AND RECALLING DATA
The PC 700 can retain up to 100 data points into memory for
later retrieval.
1) In the measurement mode, press
to insert the measured value
into memory. The stored memory location value (StO) is briefly
displayed.
2) To recall data from memory, press
. The location of the most
recent stored data is displayed first. Press or to select the
location of the desired data, then press to accept.
3) Press to return to the stored data location. Press to return
to measurement mode. To erase stored data, see Section 9.15.
11