User Guide
20
8.5 Measurement Range Selection – P3.0
Setup program P3.0 is for selecting the range of measurement.
P3.1: select measurement range and corresponding cell constant
P3.
0
RNGE
P3.
1
1000
K
10
.
Press ▲and ▼keys together (ESCAPE) at anytime, to leave the Setup mode and
return to Measurement mode.
P3.1: Select Measurement Range
Set controller to the specific range of measurement. Ensure the cell you have
connected to the controller has the same cell constant as that stated in the range.
1. Key in the password “022” as per procedure in Section 5.
2. Press ▲key until screen displays P3.0 and RNGE. Press ENTER.
3. The screen will scroll P3.1 and the selected range.
4. Press ▲or ▼keys to select appropriate range and cell.
Available ranges are as follows (Default range is highlighted):
TDS Range Cell Constant 4 mA current 20 mA current
0.00 – 10.00 ppm 0.1 0.00 ppm 10.00 ppm
0.0 – 100.0 ppm 0.1 0.0 ppm 100.0 ppm
0.0 – 100.0 ppm 1 0.0 ppm 200.0 ppm
0 – 1000 ppm
1 0 ppm 1000 ppm
0.00 – 5.00 ppt 1 0.00 ppt 5.0 ppt
0.00 – 10.00 ppt 1 0.00 ppt 10.0 ppt
0.0 – 100.0 ppt 1 0.0 ppt 100.0 ppt
0.0 – 100.0 ppt 10 0.0 ppt 100.0 ppt
5. Press ENTER to confirm (The set points 1 & 2, TDS and Temperature
calibration are reset to default values).
6. Press ▲and ▼keys together (ESCAPE) to return to P3.0. Press the ▲and
▼keys together again, to return to measurement mode.