User Guide

Operating Instructions αlpha-RES1000
21
7.6.6 Input Line Resistance Adjust
This function compensates for the line resistance of the cable to its cell.
1) Follow directions in 7.6.1 to enter Configuration mode.
2) Press ENTER key until upper display shows “0.0” and lower
display shows “L.Ad”. L.Ad = line adjuster resistance (0.0 to
100.0)
3) Press or keys to input value. Press ENTER key.
4) Proceed to 7.6.7, or return to Measurement mode by pressing
and keys simultaneously (escape).
7.6.7 Reverting to factory default settings
Use this parameter to reset all settings or calibration values only to factory default. Changing from “no”
to “YES” and pressing the ENTER key resets all settings to factory reset.
1) Follow directions in 7.6.1 to enter Configuration mode.
2) Press the ENTER key. The upper display shows “no” or “YES”,
the lower display shows “deF” (default).
3) Press the or keys to select “no” or “YES”. WARNING: Press
the ENTER key resets all settings and the calibration values
will be overwritten as a result!
4) Press the ENTER key to confirm your selection and to return to Advanced Set-up mode, or return
to Measurement mode by pressing the or keys simultaneously (escape).
7.7 Calibration (CAL) sub-function
The calibration procedure in Advanced Set-up mode is identical to procedure in Calibration mode. The
only difference is that the controller will revert back to Set-up mode (instead of Measurement mode)
after successful calibration.
7.7.1 Entering Calibration mode from Advanced Set-up mode
1) Enter Advanced Set-up mode. Push ENTER key and scroll to Advanced Set-up security code
“22”. Push ENTER key again.
2) Press or keys to scroll until upper display shows “CAL”. See section 6 for complete
calibration procedures.
0.0
L.Ad
HOLD SETUP
no
dEF
HOLD SETUP