7 Advanced Set-Up Mode
7.1 Temperature Coefficient sub-function
This sub-function allows you to select correct temperature coefficient for optimum operations. For
applications in pure water or ultra-pure water industries, select “Pur” temperature coefficient option. For
all other applications, select “Lin” temperature coefficient. Controller allows further input of temperature
coefficient values, independently for process and for calibrating solutions. Default is “Pur”.
7.1.1 Selecting Pure-water or Linear Temperature Coefficient
1) Enter Advanced set-up mode. Push ENTER key and scroll to
Advanced Set-up security code “22”. Push ENTER key. Controller
displays “tC”. Press ENTER key again. Controller displays “Pur” in
upper display and “tC” in lower display. Default is “Pur”.
2) Press ∆ or ∇ keys to select between “Pur” and “Lin” temperature
coefficients. If “Lin” tC is to be selected, press ENTER key when
controller displays “Lin”.
3) If “Lin” is selected, upper display shows “2.10%”, while lower
display show “P.tC”. This option allows input of Process
Temperature coefficient, from 0 to 10%. Default is 2.10%. Use ∆
or ∇ keys to enter required temperature coefficient value.
4) Press ENTER key. Next, enter the calibration solution
temperature coefficient value, “C.tC”, from 0 to 10%. Default is
2.10%.
5) Press ∆ or ∇ keys to input desired calibration solution
temperature coefficient.
6) Press ENTER key to accept.
7) Continue with additional Advance setup procedures or return to
Measurement mode by pressing ∆ and ∇ keys (escape)
simultaneously.