User Guide
Operating Instructions αlpha-RES1000
10
6 Calibration Mode
Calibration mode can be accessed directly from Measurement mode by pressing CAL key and entering
the Calibration security code; or from the Advanced Setup mode.
6.1 Resistivity Calibration
Calibration is always carried out in the specific range selected. The Resistivity Controller allows a one-
point calibration.
1) Enter Calibration mode. While in Measurement mode,
press CAL key and scroll to Calibration code “11”. Press
ENTER key. The upper display reads “CAL” and lower
display, “rES”.
2) Press the ENTER key. The controller displays its last set
Cell Constant (k) of the cell as a percentage of the
theoretical value. Using the up and down key, the value can
be set to anywhere from 80 to 120%. Press the ENTER key
again to carry out calibration.
3) Immerse the Conductivity cell in a suitable standard
solution, whose value is within the measurement range
selected in the controller. Agitate the Cell in the solution to
remove any trapped air-bubbles. Note: The calibration
standard should have a value that is between 20% to 100%
of the range selected. For example, if the range in the
controller is selected to be 19.99 mΩ (range 1), then the
calibration standard value should be 4.00 mΩ to 19.99 mΩ.
4) Once the reading has stabilized, use the ∆ or ∇ keys to
adjust the measured value to that of the standard solution.
Press the ENTER key to accept the value and the controller
displays the revised condition of the probe The lower display
shows the factor “FCt” while the upper display shows the
ratio of ideal to adjusted calibration value.
Note: The acceptable calibration window is ±40% of the displayed (default) value. If the
displayed is 10.00 mΩ standard, the values to which it can be adjusted is 6.00 to 14.00 mΩ. If
there is a calibration error, the controller displays “ERR” annunciator. To recalibrate repeat step
3. To exit from calibration, push both ∆ and ∇ keys (escape) to resume to Measurement mode.
5) Press ENTER key. If you entered calibration mode using CAL key, controller will return to
Measurement mode. If you entered calibration mode from Advanced Set-up mode, controller will
return to sub-function menu.
Note: When calibrating with manual temperature compensation, controller automatically changes from
preset process temperature to calibration temperature. After leaving the Calibration mode, controller
switches back to process temperature (for setting calibration temperature and process temperature, see
section 7.2).
CAL
rES
HOLD SETUP
100.0
CEL
HOLD CAL
HOLD CAL
18.20
25.0
mΩ
o
C
ATC
1.01
FCt
H
O
LD
C
AL
%
1
1
1
1