User Guide

Operating Instructions αlpha-RES1000
6
4.2 Function Groups
The main function and sub-function groups are organised in a matrix format for configuration and
selection of parameters. The main function groups are:
1) Temperature Coefficient settings (tC)
2) Temperature Measurement / compensation settings (SEt
o
C)
3) Control relay 1 configuration (SP1)
4) Control relay 2 configuration (SP2)
5) Control type (Cntr)
6) Current output (rng)
7) Configuration (ConF)
8) Calibration (CAL rES)
rng
Cntr
SETUP
HOLD
tc
SEt
SP1
SP2
rES
o
C
SETUP
HOLD
SETUPHOLD
SETUP
HOLD SETUP
HOLD
SETUP
HOLD
SETUP
HOLD
SETUP
HOLD
Set-up parameters can be viewed or changed by entering a security code. See Section 5.2 for security
code information.
4.2.1 How to view operating parameters without access to change them:
a) Press ENTER key. Display will prompt to enter security code
(S.Cd). Leave security code at “000” (do not enter a security
code).
b) Press ENTER key again. This allows for viewing (not change)
any sub-functions’ settings.
c) Press or keys to scroll through sub-functions.
d) Press ENTER key at a particular sub-function to view in detail.
e) Press ENTER key to return to sub-function menu.
f) Press and keys simultaneously (Escape key) at any time to return to Measurement mode.
Note: To simplify operations, the controller will not display parameters that are not relevant to a
particular sub-function.
000
S.Cd