Connection for panel mounting version
1. AC mains live wire 17. no connection
2. AC mains neutral wire 18. no connection
3. AC mains protective earth wire 19. 12V Power supply
4. Alarm relay (NO) 20. 12V ground
5. Alarm relay common 21. no connection
6. Alarm relay (NC) 22. Earth ground
7. Wash relay 23. no connection
8. Wash relay 24. Temperature ground
9. Relay B (SP 2) 25. Temperature input
10. Relay B (SP 2) 26. Temperature sense (short to terminal 25 if
11. Relay A (SP 1) using 2- wire RTD
12. Relay A (SP 1) 27. PMP, liquid ground in symmetrical mode
13. Hold function 28. 4-20 mA temperature output, -ve terminal
14. Hold function 29 4-20 mA temperature output, +ve terminal
15. no connection 30. 4-20 mA pH output, -ve terminal
16. no connection 31. 4-20 mA pH output, +ve terminal
IMPORTANT: The Alarm relay functions as an “Active Low” device i.e. it switches OFF under
Alarm condition. Therefore the Alarm display device should be connected to the ‘NC’ contacts
of the relay (5 & 6).
NOTE:
Switch or circuit breaker shall included in the building installation.
It shall be in close proximity to the equipment and within easy reach of the
operator.
It shall be marked as the disconnecting device for the equipment.