User Guide

Instruction Manual pH 500
- 17 -
SETUP
HOLD
SETUP
HOLD
SETUP
HOLD
6.4 HOLD Current Setting
When Transmitter is in CAL or SETUP modes, it automatically
goes into a ‘HOLD’ mode.
To indicate Transmitter is in ‘HOLD’ mode, output current can be
set to 22 mA output by activating the ‘HLD On’.
1. Press ENT key and use or keys to scroll till LCD
displays ‘SEt’ in the upper display; and ‘HLd’ in the
lower display. Press ENT again.
2. Upper display now shows ‘HLd’. Lower display will
show either ‘OFF’ or ‘On’. Use or keys to toggle
between ‘On’ of ‘OFF’. Press ENT to accept selection.
3. Press S and T keys together, to return to MEAS
mode.
NOTE: If ‘HLd’ is set to ‘OFF’, then current output will be
equivalent to last measured value.
6.5 Out-of-Range Current Setting
If pH or ORP readings exceed the scale, LCD displays ‘Or’ (over-
range), or ‘Ur’ (under-range). Transmitter current output can be
set to 3.8 mA to indicate Out-of Range situation.
1. Press ENT key and use or keys to scroll till LCD
displays ‘SEt’ in the upper display; and ‘org’ in the lower
display. Press ENT again.
2. Upper display now shows ‘org’. Lower display will show
either ‘OFF’ or ‘On’. Use or keys to toggle
between ‘On’ of ‘OFF’. Press ENT to accept selection.
3. Press S and T keys together, to return to MEAS mode.
NOTE: If ‘org’ is set to ‘OFF’, under-range will be set at 4 mA,
while over-range will be set at 20 mA.