5.4 HOLD Current Setting
When Transmitter is in CAL or SETUP modes, it automatically goes into a
‘HOLD’ mode.
To indicate Transmitter is in ‘HOLD’ mode, output current can be set to 22
mA output by activating the ‘HLD On’.
1. Press ENT key and use or keys to scroll till LCD displays
SEt’ in the upper display; and ‘HLd’ in the lower display. Press
ENT again.
2.
Upper display now shows ‘HLd. Lower display will show either
OFF’ or ‘On’. Use or keys to toggle between ‘On’ of ‘OFF’.
Press ENT to accept selection.
3. Press S and T keys together, to return to MEAS mode.
NOTE: If ‘HLd’ is set to ‘OFF’, then current output will be equivalent to last
measured value.
5.5 Out-of-Range Current Setting
If pH or ORP readings exceed the scale, LCD displays ‘Or’ (over-range), or
‘Ur’ (under-range). Transmitter current output can be set to 3.8 mA to indicate
Out-of Range situation.
1. Press ENT key and use or keys to scroll till LCD displays
SEt’ in the upper display; and ‘org’ in the lower display. Press
ENT again.
2.
Upper display now shows ‘org’. Lower display will show either ‘OFF
or ‘
On’. Use or keys to toggle between ‘On’ of ‘OFF’. Press
ENT to accept selection.
3. Press S and T keys together, to return to MEAS mode.
NOTE: If ‘org’ is set to ‘OFF’, under-range will be set at 4 mA, while over-
range will be set at 20 mA.