5 ADVANCED SETUP FUNCTIONS
The advanced setup mode lets you customize your transmitter’s preferences and
defaults. This transmitter features different sub groups that organize all setup
parameters.
The sub-groups are:
5.1 Offset Setting (Unavailable in ORP mode)
As a result of the continuous measurement and the application concerned, it may
not be convenient to remove the electrode for calibration. In such cases, an on-line
Offset adjustment is offered.
1. Press ENT key. Transmitter displays ‘OFS’. Press ENT
again. Upper display shows amount of offset, in terms of pH;
lower display shows current measured pH value.
2. Use
or keys to introduce offset to pH readings. As the
lower measured reading changes, upper display records the
amount of offset. Maximum of ± 2.00 pH offset is allowed.
3. Press both or keys together to return to ‘ MEAS ‘ mode.