Instruction manual

Instruction manual αlpha DO 2000
28
6.9 Configuration (CNFG) sub-function
In this sub-function you configure the transmitter to suit your requirements.
SETUP
SETUP
HOLD
HOLD
ENT
SETUP
SETUP
SETUP
HOLD
HOLD
HOLD
ENT
ENT
ENT
ENT
ENT
SETUP
SETUP
HOLD
HOLD
do
do
do
mg/l
ppm
SETUP
HOLD
%
SETUP
HOLD
MEAS
%
°C
ATC
ENT
ENT
ENT
SETUP
HOLD
SETUP
HOLD
SETUP
HOLD
SETUP
HOLD
EL
EL
ENT
SETUP
HOLD
ENT
ENT
UNIT
UNIT
UNIT
SETUP
HOLD
CAL
31.0
FCT
ENT
0-40
0-10
1. Select the “CNFG” subfunction, then press the ENT key.
2. Enabling/disabling the display backlight: press the or key to switch display backlight
on (BKLT ON) or off (BKLT OFF). Press the ENT key to confirm your selection.
3. Setting the backlight intensity (appears only If backlight is on): press the or key to set
backlight intensity (setting range: minimum 1 to maximum 4). Press the ENT key to confirm
your setting.
4. Selecting the unit of measurement: press the or key to select “mg/l” or “ppm” or
“%”type. Press the ENT key to confirm your selection.
5. Setting the alarm delay time in seconds: press the or key to set the alarm delay time
(setting range: 0 to 2000 seconds). Press the ENT key to confirm your setting.
Note: With this parameter you set the delay before the alarm relay activates when
the set point has been exceeded.