Instruction manual

Instruction manual αlpha DO 2000
27
6.8 Wash relay (WASH) sub-function
In this sub-function you set the parameters for the wash relay.
SETUP
SETUP
HOLD
HOLD
ENT
SETUP
SETUP SETUP
HOLD HOLD HOLD
ENT
ENT ENT
ENT
ENT
SETUP
SETUP
HOLD
HOLD
1. Select the “WASH” sub-function, then press the ENT key.
2. Enabling/disabling wash function: press the or key to enable (WASH ON) or disable
(WASH OFF) wash function. Press the ENT key to confirm your selection.
3. Selecting the relay status condition: press the or key to choose desired relay status
(de-energized = “DEEN” or energized = “EN”). Press the ENT key to confirm your selection.
4. Setting the wash interval in hours: Press the or key to set the desired wash interval
(setting range: 0.1 to 200.0 hours). Press the ENT key to confirm your setting.
5. Setting the wash duration in seconds: Press the or key to set the desired wash
duration (setting range: 1 to 2000 seconds). Press the ENT key to confirm your setting.
Note: During wash cycle the transmitter is set to HOLD. For safety reasons, the
HOLD function is activated 5 seconds prior and deactivated 10 seconds after the
wash cycle.
Continue with Setup mode procedures, or return to Measurement mode by pressing the
and keys simultaneously (escape).