Instruction manual

Instruction manual αlpha DO 2000
26
6.7 Current Output 2 sub-function
In this sub-function you set the current output range of the transmitter for temperature values.
1. Select the “CUR.2” sub-function, then press the ENT key.
2. Selecting the output type: press the or key to select the desired output type: 0-20 mA
or 4-20 mA. Press the ENT key to confirm your selection.
3. Setting the temperature value at which the transmitter output will be 4 mA (or 0 mA):
press the or key to set the temperature value (setting range: -10.0 to 115.0 °C or 14 to
237 °F) to be equivalent to 4 mA (or 0 mA). Press the ENT key to confirm your setting.
4. Setting the temperature value at which the transmitter output will be 20 mA: press the
or key to set the temperature value (setting range: 0.0 to 125.0 °C or 32 to 257 °F) to be
equivalent to 20 mA. Press the ENT key to confirm your setting.
Continue with Setup mode procedures, or return to Measurement mode by pressing the
and keys simultaneously (escape).
SETUP
HOLD
SETUP
HOLD
ENT
ENT
SETUP
HOLD
ENT
SETUP
SETUP
HOLD HOLD
°C °C
°C °C
ENT
ENT
ENT
ENT
SETUP SETUP
HOLD
HOLD