Instruction manual
Instruction manual αlpha DO 2000
25
6.6 Current Output 1 sub-function
In this subfunction you set the current output range of the transmitter for dissolved oxygen
values.
90.0
90.0
ENT
SETUP
HOLD
SETUP
HOLD
ENT
ENT
SETUP
HOLD
ENT
SETUP
SETUP
HOLD
HOLD
%
%
SETUP
SETUP
HOLD
HOLD
ENT
ENT
ENT
ENT
SETUP
SETUP
HOLD
HOLD
%
%
1. Select the “CUR.1” subfunction, then press the ENT key.
2. Selecting the output type: press the ▲ or ▼ key to select the desired output type: 0-20 mA
or 4-20 mA. Press the ENT key to confirm your selection.
3. Setting the dissolved oxygen value at which the transmitter output will be 4 mA (or 0 mA):
press the ▲ or ▼ key to set the dissolved oxygen value to be equivalent to 4 mA or 0 mA,
respectively. Press the ENT key to confirm your setting.
4. Setting the dissolved oxygen value at which the transmitter output will be 20 mA: press
the ▲ or ▼ key to set the dissolved oxygen value to be equivalent to 20 mA. Press the ENT
key to confirm your setting.
5. Selecting out of range current: Press the ▲ or ▼ key to switch “CU22” on or off. If “CU22”
is on, the output signal will jump to 22 mA when the measured dissolved oxygen value is below
or above the values set in points 3 & 4 above. Press the ENT key to confirm your setting.
Note: the “CU22” setting (on/off) of current output 1 applies to current output 2 as
well.
Continue with Setup mode procedures, or return to Measurement mode by pressing the ▲
and ▼ keys simultaneously (escape).