6.4 Control Relay A / Relay B (SP1/SP2) sub-function
The SP1 sub-function determines the operating parameters for Relay A; while SP2 defines the
operating parameters for Relay B. Since these groups have the same set-up parameters, they
are described together.
SETUP
SETUP
HOLD
HOLD
ENT
SETUP
SETUP
HOLD
HOLD
ENT
SETUP
SETUP
SETUP
HOLD
HOLD
HOLD
ENT
ENT
ENT
%
%
90.0
2.0
1. Select the “SP1” (Relay A) or “SP2” (Relay B) sub-function, then press the ENT key.
2. Setting set point value: press the or key to enter the value for set point 1 (set point 2)
at which your controller will activate. Press the ENT key to confirm your setting.
3. Selecting relay function: press the or key to select the desired relay function (“LO”=
low or “HI”= high). Press the ENT key to confirm your selection.
Note: Note: This parameter lets you choose the relay function. Select “LO” to
activate the relay when the dissolved oxygen value undershoots the set point;
select “HI” to activate the relay when the value overshoots the set point. SP1 and
SP2 can be selected as “Lo/Lo”, “Lo/Hi”, “Hi/Lo”, or “Hi/Hi”.
4. Setting a hysteresis value: press the or key to select the desired hysteresis (setting
range: 1 to 10% or 0.1 to 1.0 mg/l) for set point 1 (set point 2). Press the ENT key to confirm
your setting.
Note: Hysteresis prevents rapid contact switching if your value is fluctuating near
the set point. Please refer to Appendix 3 for a graphical representation of the
hysteresis.
Example: You have set your high set point at 90 % and your hysteresis value is 2 %. If your
measured value overshoots 90 %, the controller’s relay activates. The actions of the external
device will cause the solution’s dissolved oxygen percentage saturation to drop. The relay will
deactivate, when the dissolved oxygen percentage saturation value drops below 88 %.