19
6.3 Setting temperature (Set °C°F) sub-function
1. Select the “SET °C°F” sub-function, then press the ENT key.
2. Selecting temperature unit: press the ▲ or ▼ key to select the desired temperature unit
“°C” or “°F”. Press the ENTER key to confirm your selection.
3. Enable/disable ATC mode: press the ▲ or ▼ key to enable (ATC ON) or disable (ATC OFF)
automatic temperature compensation. Press the ENT key to confirm your selection.
− ATC disabled (ATC OFF):
For manual temperature compensation you can set the process and calibration
temperatures. This allows calibration at a temperature other than the process
temperature. Example: setting a calibration temperature of 25°C lets you calibrate
using standard solutions at 25°C, even if your process temperature is different from
25°C.
4. Setting the process temperature: press the ▲ or ▼ key to adjust the process temperature
(upper value, setting range: –10.0 to 125.0°C / 14.0 to 257.0 °F). Press the ENT key to confirm
process temperature.
5. Setting the calibration temperature: press the ▲ or ▼ key to adjust the calibration
temperature (upper value, setting range: – 10.0 to 125.0°C / 14.0 to 257.0 °F). Press the ENT
key to confirm the calibration temperature.
− ATC enabled (ATC ON):
6. Setting the temperature offset: The display shows the current temperature offset (upper value)
and the temperature currently measured (lower value). Compare the current temperature
displayed on the transmitter to a thermometer known to be accurate. Note the correct
temperature value. Press the ▲ or ▼ key to adjust the lower value. The upper number
indicates the offset value. You can offset the temperature by up to ±10°C/±18°F. Press the
ENT key to confirm your setting.
Continue with additional Setup procedures, or return to the Measurement mode by pressing
the ▲ and ▼ key (escape) simultaneously.