User Guide

Operating Manual alpha DO 1000
26
7.7.4 Selecting Dissolved Oxygen value at 20mA
This parameter lets you choose the Dissolved Oxygen value at which the
transmitter output will be 20mA.
1. Follow directions in 7.7.1 to enter Current Output mode.
2. Press the ENTER key until the upper display shows a Dissolved Oxygen
value and the lower display shows “r.20”.
3. Press the or keys to select dissolved oxygen value to be equivalent
to 20 mA (Default is 8.00mg/l or 100%).
4. Press the ENTER key to confirm your selection and return to Advanced
Set-up mode, or return to Measurement mode by pressing the and
keys simultaneously (escape).
7.8 Configuration (ConF) sub-function
This group of parameters lets you configure the controller to suit your requirements.
HOLD HOLDHOLD
mg/l
HOLD
HOLD
HOLD
SETUP SETUPSETUP SETUP
SETUP
SETUP
C F0
u
S Y
0-40
Al.o Al.C
el
EF
7.8.1 Entering the Configuration sub-function
1. Enter Advanced Set-up mode. Push the ENTER key and scroll to
Advanced Set-up security code “22”. Push the ENTER key again.
2. Press the or keys to scroll until the upper display shows “ConF”.