7.4.3 Setting Manual Temperature Compensation
Note: This parameter is blanked out when the controller is set for
ATC operation.
For manual temperature compensation, you can set two different temperatures:
process and calibration. This allows calibration at a temperature other than your
process temperature. Example: setting a calibration temperature of 25 °C lets you
calibrate using standard buffer solutions at 25 °C, even if your process temperature
is a different temperature.
1. Select “ATC oFF” as described above in section 7.4.1
2. Press the ENTER key. The upper display shows the
current process temperature and the lower display
shows “P.°C” to indicate process temperature.
3. Press the or keys to adjust the process
temperature value. You can adjust the value from –
9.9 to 125 °C.
4. After you set the process temperature value, press
the ENTER key. The upper display shows the current
calibration temperature and the lower display shows
“C.°C” to indicate calibration temperature.
5. Press the or keys to adjust the calibration
temperature value. You can adjust this value from –
9.9 to 125 °C.
6. Press the ENTER key to confirm your selection.
7. Continue with additional Advanced Set-up procedures,
or return to Measurement mode by pressing the or keys (escape)
simultaneously.
7.5 Control Relay A/Control Relay B (SP1/SP2) sub-function
The SP1 option sets the operating parameters for Relay A; and SP2 for relay B.
Since these groups have the same set-up parameters, they are described together.
Sp1 Sp1 Hys O . OF.
mg/l mg/l
HOLD HOLD HOLD HOLD HOLD HOLDSETUP SETUP SETUP SETUP SETUP SETUP
Sp1 6.20 L 0.20 0 0