User Guide
Operating Manual alpha DO 1000
15
7 ADVANCED SET-UP MODE
7.1 Electrode Offest (OFS), Salinity and Pressure sub-function
This mode allows you to change the offset parameter to make reading corrections
without removing the electrode from the control system. You can make adjustments
of up to ±2.00ppm; ±2.00 mg/l; or ±10.0%. At the end of the offset adjustment, the
salinity value, followed by the pressure values (in mmHg or Pascal) respectively, is
entered.
HOLD
HOLD
HOLD
HOLD
HOLD
SETUP
SETUP
SETUP
SETUP
SETUP
mg/l ppt
ofs 0.02
760
0.0
8.24 sal
1. In this mode, the controller will add or subtract the value from the
measured dissolved oxygen value and display the correct value. However,
if you need to offset the value beyond the average offset you would expect
in your application type, consider a full calibration or even electrode
replacement.
2. Take a sample from the system. Record the controller’s Dissolved Oxygen
reading at the time the sample was taken.
3. Measure the Dissolved Oxygen of your sample using a calibrated
Dissolved Oxygen hand-held meter or bench meter (For a list of Eutech
Instruments’ DO hand-held meter, refer to section on Accessories).
4. Record the correct Dissolved Oxygen value.
5. Enter Advanced set-up mode. Push the ENTER key and scroll to
Advanced Set-up security code “22”. Push the ENTER key again.
6. Press the ▲ or ▼ keys to scroll until the upper display shows “OFS”.
7. Press the ENTER key. The SETUP annunciator appears at the top of the
display. The lower display shows the current measured Dissolved Oxygen
with its respective units. The upper display shows the current offset value.
8. Press the ▲ or ▼ keys until the Dissolved Oxygen value coincides with
the correct Dissolved Oxygen value noted in Step 2. As the ▲ or ▼ keys is