User Guide
Operating Manual alpha DO 1000
14
annunciator comes on. Press ENTER key to accept the value and the
controller displays the status of the probe.
4. For Two-Point Calibration, calibration is carried out at 0% saturation first.
Immerse the probe in solution and allow it to stabilise. When the reading is
stable, the “READY” annunciator comes on. Press the ENTER key to
accept the value.
Note: If the displayed value is > ±5% of the theoretical value, the
controller will not accept the zero calibration and retains the
previous calibration values.
5. If the calibration is successful, the controller switches to the second point,
100% saturation. Take the probe out of the zero solution, rinse in clean
water and calibrate in air. When the reading is stable, “READY”
annunciator comes on. Press ENTER key to accept the value and the
controller displays the status of the probe.
Note: If the displayed value is > ±30% of the theoretical value, the
controller will not accept the high-level calibration and retains the
previous calibration values.
6. After calibration to a second value, this controller will automatically display
condition of the probe in the upper display and the temperature in the
lower display. You can view these values without calibration --- see section
5.1.1 for instruction. The probe condition is re-determined after each
calibration.
7. Press the ENTER key. If you entered the calibration mode using the CAL
key, the controller will return to the Measurement mode. If you entered the
calibration mode from the Advanced Set-up mode, the controller will return
to the sub-function menu.
Note: If there is a calibration error, the controller displays “ERR”. If
this happens, push both the ▲ and ▼ keys (escape) to restart
the calibration beginning from step 1.
Note: When calibrating with manual temperature compensation, the
controller automatically changes from the preset process
temperature to the calibration temperature. After leaving the
Calibration mode, the controller switches back to process
temperature (for setting the calibration temperature and the
process temperature, see section 7.4.3).